XENON xenon lamp UV/Vis spectrophotometer — double beam, 190–1100 nm
The XENON xenon lamp UV/Vis spectrophotometer combines double beam optics with a Hamamatsu xenon flash lamp. The lamp needs no warm-up, so the instrument measures as soon as it is switched on, and one source covers 190–1100 nm. Two models are available — XENON-200 with a spectral bandwidth of 2 nm and XENON-250 with 1 nm — both with scanning up to 3000 nm/min and a 10.1″ touch screen.
Which UV/Vis spectrophotometer?
Advanced — double beam, variable bandwidth, PMT, xenon
Double beam optics with a xenon flash lamp
Deuterium and tungsten lamps burn continuously and need time to stabilise before the first reading. A xenon lamp UV/Vis spectrophotometer works differently: the flash lamp is fired only during the measurement and reaches full intensity at once. In XENON the reference beam compensates pulse-to-pulse variations, so the baseline stays stable to ±0.0003 A per hour, with stray light ≤0.03 %T at 220 and 360 nm and noise of 0.0005 A at 500 nm.

Key characteristics
- Double beam UV/Vis spectrophotometer, grating 1200 lines/mm, wavelength range 190–1100 nm
- Hamamatsu xenon flash lamp — no warm-up; according to the manufacturer the lamp lasts at least 3 years
- Two bandwidth versions: XENON-200 2 nm (baseline ±0.002 A) · XENON-250 1 nm (baseline ±0.001 A)
- Wavelength accuracy ±0.3 nm, repeatability ≤0.1 nm; stray light ≤0.03 %T; stability ±0.0003 A/h
- Photometric accuracy ±0.002 A (0–0.5 A); photometric range −0.3 to 3 A; noise 0.0005 A at 500 nm
- Scanning up to 3000 nm/min, kinetics, quantitation, multi-wavelength and DNA / protein test; 10.1″ touch screen, USB and RS232
Suitable for
- Laboratories with occasional or interrupted measurements that should not wait for a lamp to warm up
- Pharmaceutical and quality control laboratories with UV/Vis methods and spectral scans
- Biochemistry and life sciences: nucleic acids and proteins at 260/280 nm, enzyme kinetics
- Environmental and water laboratories with UV methods such as nitrate
- Teaching and research laboratories that need one instrument for the whole UV and visible range
Why a xenon lamp UV/Vis spectrophotometer
A xenon flash lamp emits short, intense light pulses across the whole UV and visible range. Because it is fired only for the measurement, there is no warm-up period and no change-over between a deuterium and a tungsten lamp — the spectrum from 190 to 1100 nm comes from one source. The lamp works only while the instrument measures, which is why the manufacturer states a lamp life of at least 3 years.
For continuous routine work with deuterium and tungsten lamps choose the LEGEND or NOVEL-900; for the highest sensitivity in the UV choose the LEGEND-PMT with a photomultiplier detector — all in the menu above.
Cell holders and sample handling
The large sample compartment of the XENON xenon lamp UV/Vis spectrophotometer accepts cuvettes with path lengths from 5 to 100 mm. A single 10 mm cell holder and six cuvettes are in standard delivery; automatic 8-cell holder, film holders, integrating sphere, reflection accessory and a Peltier / sipper system are available as options.
XENON xenon lamp UV/Vis spectrophotometer — technical specifications
Optics and photometry
| Optical system | Double beam; grating 1200 lines/mm |
| Wavelength range | 190–1100 nm |
| Spectral bandwidth | XENON-200: 2 nm · XENON-250: 1 nm |
| Wavelength accuracy | ±0.3 nm |
| Wavelength repeatability | ≤0.1 nm |
| Photometric accuracy | 0.2 %T; ±0.002 A (0–0.5 A); ±0.004 A (0.5–1 A) |
| Photometric repeatability | ≤0.15 %T; 0.001 A (0–0.5 A); 0.002 A (0.5–1 A) |
| Photometric range | 0–200 %T; −0.3 to 3 A; 0–9999 C |
| Stray light | ≤0.03 %T at 220 and 360 nm |
| Stability | ±0.0003 A/h at 500 nm |
| Baseline flatness | XENON-200: ±0.002 A · XENON-250: ±0.001 A |
| Noise | 0.0005 A at 500 nm |
| Detector | Silicon photodiodes |
Functions, data and general
| Measuring modes | T, A, C, E |
| Functions | Photometry, quantitation, kinetics, wavelength scan, multi-wavelength, DNA / protein |
| Scanning speed | High, medium, low · max. 3000 nm/min |
| Light source | Hamamatsu xenon flash lamp, no warm-up |
| Cell holder | Single 10 mm (standard) · path length 5–100 mm |
| Display | 10.1″ capacitive touch screen |
| Data memory | Up to 1024 MB |
| Data output | USB drive, USB host, RS232 |
| PC software | Included |
| Power supply | AC 110–220 V, 50–60 Hz |
| Packing size | 810 × 660 × 390 mm |
Operation and data

Photometry, wavelength scan, quantitation, multi-wavelength measurement, kinetics and the DNA / protein test are started directly from the 10.1″ touch screen. The wavelength is set automatically; the scanning speed is selectable up to 3000 nm/min.
Up to 1024 MB of test data and working curves stay in the instrument; results go to a USB drive, to a PC over USB or RS232. PC software is in standard delivery; 21 CFR compliant software and a built-in printer are options.
Standard delivery
- XENON UV/Vis double beam spectrophotometer with xenon flash lamp (XENON-200 or XENON-250)
- Single cell holder, 10 mm
- 4 × glass cuvette and 2 × quartz cuvette, 10 mm path length
- PC software, power cable and user manual
Optional: automatic 8-cell holder, manual 4-cell holder up to 100 mm, micro cell holder, film holders, integrating sphere, reflection accessory, Peltier / sipper system, test-tube rack, built-in printer and 21 CFR compliant software. The delivered configuration for a specific order is the one confirmed in the COLO.Science quotation.
Knowledge base
Spectral bandwidth, stray light, Beer–Lambert law and the choice of cuvettes are explained on COLO.Science — UV-Vis spectrophotometry and colour measurement, our knowledge base; how absorbance fits among other laboratory measurements is described in Measurement in laboratory science. The full data are in the XENON technical specification (PDF).
Selection guide
- Measurement without warm-up, one lamp for 190–1100 nm: XENON xenon lamp UV/Vis spectrophotometer — double beam, 2 or 1 nm, 10.1″ touch screen.
- Double beam stability at an entry price: NOVEL-102S or NOVEL-800 — 190–1100 nm, 2 nm, 8″ touch screen.
- Six models, single or double beam, 7″ colour screen: LEGEND series — bandwidth 2, 1 or 0.5–5 nm.
- Selectable bandwidth and the lowest stray light: NOVEL-900 double beam with 0.5–5 nm.
- GLP data management: MAVRICA-T series — 11″ touch screen, cloud and wireless data output.
- Highest sensitivity in the UV: LEGEND-PMT double beam with PMT detector — 190–900 nm.
Configuration notice
This page is provided for orientation and product selection. Cell holders, cuvettes, printer, software and other options are confirmed with the official COLO.Science quotation.
Sales and support for international customers
COLO Lab ExpertsPolje ob Sotli 4, SI-3255 Buče, Slovenia
Phone: +386 70 713 437 · Email: sales@colo.si · Web: colo.si
We supply laboratories across Europe directly and through partners. B2B prices are quoted net, excluding VAT.