COLO.Science · Spectroscopy & Color · UV/Vis spectrophotometers
NOVEL4S UV/Vis scanning spectrophotometer — 190–1100 nm, spectral bandwidth 2 nm
The NOVEL4S UV/Vis scanning spectrophotometer is a single-beam instrument for qualitative and quantitative analysis across the ultraviolet and visible range. An ARM processing core with automatic wavelength setting, a 1200 lines/mm grating with 2 nm spectral bandwidth and deuterium plus tungsten-halogen lamps cover 190–1100 nm. Photometric, quantitative, scanning, kinetics and DNA/protein methods run directly on the 7″ colour touch screen.
Which UV/Vis spectrophotometer?
Single beam — routine UV/Vis
Double beam, variable bandwidth and xenon
Full UV/Vis range for routine and method work
Most routine UV/Vis methods — nucleic acid and protein checks at 260/280 nm, nitrate and phosphate in water, colour reactions in the visible range — are written for a bandwidth of 2 nm or less. The NOVEL4S UV/Vis scanning spectrophotometer, a 2 nm bandwidth UV/Vis spectrophotometer, meets that with single-beam optics that stay simple to operate, and adds full spectral scanning so an unknown sample can be checked before a quantitative method is set up.

Key characteristics
- Wavelength range 190–1100 nm with deuterium and tungsten-halogen lamps
- Spectral bandwidth 2 nm, single beam, 1200 lines/mm grating, 170 mm focal length
- Wavelength accuracy ±0.3 nm, repeatability ≤0.1 nm, automatic wavelength setting
- Photometric accuracy ±0.002 A (0–0.5 A); stray light ≤0.03 %T at 360 nm
- Full-band and sub-band scanning, kinetics up to 999 min, DNA/protein mode
- 7″ colour touch screen, USB data output; built-in printer and PC software as options
Suitable for
- Medicine, health and clinical testing laboratories
- Biochemistry and life science: a UV/Vis spectrophotometer for DNA/protein and kinetics, plus colorimetric assays
- Environmental protection and water analysis with photometric methods
- Petrochemical laboratories and quality control of raw materials and products
- Teaching and university laboratories that need the full UV range
Why a 2 nm single beam scanning spectrophotometer
A narrower slit separates neighbouring absorption bands better, a wider slit lets more light through. At 2 nm the NOVEL4S UV/Vis scanning spectrophotometer sits where most pharmacopoeia-style and standard methods expect it, so results can be compared with published methods without correction. Single beam scanning spectrophotometer optics keep the instrument compact and affordable; automatic zero, power-off memory of calibration curves and automatic light door protection of the detector keep daily work quick and repeatable.
When the method needs a selectable bandwidth, very long measurement sequences or no lamp warm-up, compare the double-beam and xenon models in the menu above.
Cell holders and sample handling
The NOVEL4S UV/Vis scanning spectrophotometer is delivered with a manual 4-position holder for 10 mm cuvettes. The sample compartment accepts holders from micro cells up to 100 mm path length, film holders, an integrating sphere and a Peltier / sipper system, so the same instrument can follow the samples as the work changes.
NOVEL4S UV/Vis scanning spectrophotometer — technical specifications
Optics and photometry
| Optical system | Single beam, grating 1200 lines/mm |
| Focal length | 170 mm |
| Wavelength range | 190–1100 nm |
| Spectral bandwidth | 2 nm |
| Wavelength accuracy | ±0.3 nm |
| Wavelength repeatability | ≤0.1 nm |
| Photometric accuracy | ±0.2 %T (0–100 %T); ±0.002 A (0–0.5 A); ±0.004 A (0.5–1 A) |
| Photometric repeatability | ≤0.1 %T; 0.001 A (0–0.5 A); 0.002 A (0.5–1 A) |
| Photometric range | 0–200 %T; −3 to 3 A; 0–9999 C |
| Stray light | ≤0.03 %T at 360 nm |
| Stability | ≤±0.0005 A/h at 500 nm |
| Noise · baseline drift | 0.0005 A at 500 nm · 0.003 A / 0.5 h |
| Light source · detector | D2 and tungsten-halogen lamps · silicon photodiode |
Functions, data and general
| Measuring modes | T, A, C, E; photometric, quantitative, spectral scanning, kinetics, DNA/protein |
| Quantitative analysis | Up to 10 calibration standards; curves with up to 3 wavelengths |
| Scanning interval · speed | 0.1 / 0.2 / 0.5 / 1 / 2 / 5 nm · fast, medium, slow |
| Kinetics | Up to 999 min, interval 0.1–30 s |
| Units | None, %, ppb, mol/L, g/L, mg/mL, µg/mL, ng/mL |
| Display | 7″ colour LCD touch screen |
| Data output | USB |
| Printer · PC software | Built-in thermal printer · COLO spectrum software (options) |
| Power supply | AC 220 V ±10 %, 50 Hz, 110 W |
| Packing size · gross weight | 630 × 510 × 500 mm · 25 kg |
Operation and data

The NOVEL4S is a 7 inch touch screen UV/Vis spectrophotometer: wavelength, method and units are set on the screen; spectra, calibration curves and kinetic plots are shown while the measurement runs. Concentration is read directly from a factor or from a regression curve with up to 10 standards.
Test data, spectra, regression equations and instrument correction parameters are kept after power-off. Results are exported via USB; the built-in thermal printer and the COLO spectrum PC software are available as options for reports and further data processing.
Standard delivery
- NOVEL4S UV/Vis scanning spectrophotometer with 7″ touch screen
- Manual 4-position cuvette holder, 10 mm
- 4 × glass cuvette and 2 × quartz cuvette, 10 mm path length
- Power cable with Euro plug and instruction manual
Optional: built-in thermal printer, COLO spectrum PC software, Wi-Fi UV lamp control, automatic 8-cell holder, cell holders up to 100 mm, micro cell holder, film holders, integrating sphere, flow cell, Peltier / sipper system and test-tube rack. The delivered configuration for a specific order is the one confirmed in the COLO.Science quotation.
Knowledge base
Spectral bandwidth, stray light, Beer–Lambert law and the choice of cuvettes are explained on COLO.Science — UV-Vis spectrophotometry and colour measurement, our knowledge base; how absorbance fits among other laboratory measurements is described in Measurement in laboratory science. The full data are in the NOVEL4S technical specification (PDF).
Selection guide
- Routine UV/Vis at 2 nm with scanning: NOVEL4S UV/Vis scanning spectrophotometer — full range, touch-screen operation, compact single-beam optics.
- Several bandwidths in one single-beam instrument: MAVRICA18 or LEGEND series — bandwidth from 0.5 to 5 nm.
- Long sequences, baseline stability, resolution checks: NOVEL-900 double beam with selectable 0.5–5 nm.
- GLP data management and wireless output: MAVRICA-T double and split beam series.
- No warm-up, intermittent or high-throughput work: XENON-PMT with xenon lamp and PMT detector.
Configuration notice
This page is provided for orientation and product selection. Cell holders, cuvettes, printer, software and other options are confirmed with the official COLO.Science quotation.
Sales and support for international customers
COLO Lab ExpertsPolje ob Sotli 4, SI-3255 Buče, Slovenia
Phone: +386 70 713 437 · Email: sales@colo.si · Web: colo.si
We supply laboratories across Europe directly and through partners. B2B prices are quoted net, excluding VAT.