COLO.Science · Spectroscopy & Color · UV/Vis spectrophotometers
MAVRICA-M UV/Vis spectrophotometer with Wi-Fi — five models, 11″ touch screen
The MAVRICA-M UV/Vis spectrophotometer with Wi-Fi is a series of five models on one platform with an 11″ touch screen: single beam M50, M51 and M51S for 190–1100 nm with 2 nm, 1 nm or a selectable 0.5–5 nm, M30 for the visible range 320–1100 nm and the split beam MS40. Results go over Wi-Fi, Bluetooth, e-mail or SMS, and a printer is connected without a cable.
How the MAVRICA-M measures and shares results
The single beam models measure the blank and the sample one after the other in the same light path; the split beam MS40 also sends part of the light to a reference detector, which follows the lamp intensity. Results are shown on the 11″ touch screen and leave the MAVRICA-M UV/Vis spectrophotometer with Wi-Fi, Bluetooth 3.0, e-mail or SMS — without a PC next to the instrument.

Key characteristics
- Five models on one platform with an 11″ touch screen, grating 1200 lines/mm, stray light ≤0.03 %T
- Single beam: M30 320–1100 nm, 2 nm · M50 190–1100 nm, 2 nm · M51 1 nm · M51S 0.5 / 1 / 2 / 4 / 5 nm
- Split beam: MS40 190–1100 nm, 2 nm, with a reference detector
- Wavelength accuracy ±0.3 nm; photometric accuracy ±0.2 %T; range −4 to 4 A
- Photometry, quantitation, kinetics, multi-wavelength, DNA / protein and scanning up to 8000 nm/min
- Wi-Fi, Bluetooth 3.0, e-mail and SMS; GLP/GMP functions; manual 4-cell holder in standard delivery
Suitable for
- Water and environmental laboratories: nitrate, phosphate, COD and other photometric tests
- Food and beverage quality control
- Laboratories that send results to a server, by e-mail or to a phone
- Biochemistry and clinical laboratories: DNA and protein at 260/280 nm
- Teaching laboratories at schools and universities
Why a UV/Vis spectrophotometer with Wi-Fi
A UV/Vis spectrophotometer with Wi-Fi sends results where they are needed — to a printer, a server, a mailbox or a phone — without a PC and cables at the bench. The five MAVRICA-M models share the screen, the connectivity and the accessories; they differ in range, bandwidth and optics, so the model is chosen for the methods in use: M30 for visible methods, M50 for routine UV/Vis, M51 or M51S for narrower or selectable bandwidth, MS40 for longer series.
For long kinetics and scan series choose the double beam MAVRICA-T on the same platform; for a 7″ touch screen the NOVEL4S, for an automatic 8-cell holder the NOVEL-100S — both in the menu above.
Cell holders and sample handling
A manual 4-cell holder for 10 mm cuvettes, 4 glass cuvettes and 2 quartz cuvettes for UV are in standard delivery of the MAVRICA-M. An automatic 4-, 5- or 8-cell holder, film holder, test-tube rack, Peltier / sipper system, integrating sphere and reflection accessory are available as options.
MAVRICA-M UV/Vis spectrophotometer — technical specifications
Optics and photometry
| Optical system | M30, M50, M51, M51S: single beam · MS40: split beam; grating 1200 lines/mm |
| Wavelength range | 190–1100 nm · M30: 320–1100 nm |
| Spectral bandwidth | M30, MS40, M50: 2 nm · M51: 1 nm · M51S: 0.5 / 1 / 2 / 4 / 5 nm |
| Wavelength accuracy | ±0.3 nm (±0.1 nm at 656.1 nm) |
| Wavelength repeatability | ≤0.1 nm |
| Photometric accuracy | ±0.2 %T · ±0.0005 A (0–0.5 A) · ±0.0008 A (0.5–1 A) |
| Photometric repeatability | ≤0.15 %T · 0.0004 A (0–0.5 A) · 0.0006 A (0.5–1 A) |
| Photometric range | 0–400 %T; −4 to 4 A |
| Resolution | 0.00001 A |
| Stray light | ≤0.03 %T at 220 and 360 nm |
| Stability | ±0.0001 A |
| Baseline flatness | ±0.0004 A |
| Noise | 0.0005 A at 500 nm |
Functions, data and general
| Measuring modes | T, A, C, E |
| Functions | Photometry, quantitation, kinetics, wavelength scan, multi-wavelength, DNA / protein |
| Scanning speed | High, medium, low · max. 8000 nm/min |
| Light source | Deuterium and tungsten lamp · M30: tungsten lamp |
| Detector | Silicon photodiode |
| Cell holder | Manual 4-cell holder 10 mm (standard) |
| Display | 11″ touch screen |
| Data output | Cloud connection, e-mail, SMS, Bluetooth 3.0, Wi-Fi |
| Printing | Laser printer over Wi-Fi · thermal printer over Bluetooth |
| Data control | GLP / GMP: users, audit trail, log |
| Power supply | AC 110–220 V, 50–60 Hz |
Operation and data

Photometry, quantitation, kinetics, wavelength scan, multi-wavelength measurement and DNA / protein analysis are started from the 11″ touch screen. The scanning speed is high, medium or low, up to 8000 nm/min.
Results are sent over Wi-Fi or Bluetooth 3.0 (about 10 m), by e-mail or SMS, and printed on a laser printer over Wi-Fi or a thermal printer over Bluetooth. User management, audit trail and log are on the instrument; 21 CFR compliant software is available as an option.
Standard delivery
- MAVRICA-M UV/Vis spectrophotometer (M30, MS40, M50, M51 or M51S)
- Manual 4-cell holder, 10 mm
- 4 × glass cuvette and 2 × quartz cuvette for UV, 10 mm
- Power cable and user manual
Optional: automatic 4-, 5- or 8-cell holder, film holder, test-tube rack, Peltier / sipper system, integrating sphere, reflection accessory and 21 CFR compliant software. The delivered configuration for a specific order is the one confirmed in the COLO.Science quotation.
Knowledge base
Spectral bandwidth, stray light, Beer–Lambert law and the choice of cuvettes are explained on COLO.Science — UV-Vis spectrophotometry and colour measurement, our knowledge base; how absorbance fits among other laboratory measurements is described in Measurement in laboratory science. The full data are in the MAVRICA-M technical specification (PDF).
Selection guide
- Five models with Wi-Fi and 11″ touch screen: MAVRICA-M UV/Vis spectrophotometer — single or split beam.
- Same platform, double beam: MAVRICA-T — three models with GLP/GMP data control.
- Single beam with LCD and PC software: MAVRICA18 — four models.
- Automatic 8-cell holder as standard: NOVEL-100S — split beam, 8″ touch screen.
- Visible range only, basic and low-priced: MAVRICA-V or NOVEL2 — 325–1000 nm.
Configuration notice
This page is provided for orientation and product selection. Cell holders, cuvettes, printer, software and other options are confirmed with the official COLO.Science quotation.
Sales and support for international customers
COLO Lab ExpertsPolje ob Sotli 4, SI-3255 Buče, Slovenia
Phone: +386 70 713 437 · Email: sales@colo.si · Web: colo.si
We supply laboratories across Europe directly and through partners. B2B prices are quoted net, excluding VAT.