Overview — VIKING-M510T-A portable multi-parameter meter
VIKING-M510 from COLO.Science is a rugged IP65 portable instrument for pH, mV/ORP, conductivity, resistivity, TDS, salinity, ISE, dissolved oxygen and temperature. It features a bright 4.3″ color touchscreen, an intelligent OS with user/calibration/electrode/method/data management, and GLP-compliant memory (1000 groups per parameter) for traceable results. Multi-reading modes (Auto-Read, Timed-Read, Continuous) and automatic/manual temperature compensation ensure reliable measurements in the lab or in the field.
Measurement modes & calibration
- Reading modes: Auto Read (Fast/Medium/Slow), Timed, Continuous
- pH: up to 8 points; USA/NIST/DIN/GB/MERK/JIS buffers; user-defined buffers; auto slope/offset diagnostics
- Conductivity: up to 5 points; ref. temp 5/10/15/18/20/25 °C; 10 μS/cm–12.88 mS/cm (and equivalents) standard recognition
- ISE: Direct, Standard Addition, Sample Addition, GRAN; >10 built-in methods (e.g., F−, Cl−, NO3−, NH4+, K+, Na+, Ca2+, Cu2+, Pb2+, Ag+); up to 8 points
- DO: polarographic sensor; air-saturated water or zero O2 calibration; auto barometric compensation; manual salinity correction
Key features
- 4.3″ high-contrast color touchscreen; backlight
- Intelligent OS: user / calibration / electrode / method / data / log management
- Auto-Hold endpoint detection; ATC/MTC
- USB 2.0 (flash drive / PC / printer-scanner)
- Reset to factory defaults; auto power-off (300–3600 s)
- IP65 enclosure for field and outdoor measurements
Inputs & outputs
- Inputs: pH electrode BNC (Q9); DO probe 4‑pin aviation; Conductivity probe 5‑pin aviation
- Outputs: USB 2.0 (flash drive, PC, scanner/printer)
Models & dimensions
| Model | Name | Display | IP Rating | Dimensions (W×L×H) | Weight | 
| VIKING-M510T-A | Portable multi-parameter meter | 4.3″ color touchscreen | IP65 | 90 × 255 × 40 mm | ≈ 500 g | 
Technical specifications
Technical specifications — VIKING-M510T-A
| Model | M510T | 
| Parameters | pH / EC / ISE / DO / Temp. (mV / ORP / pX / Resistivity / TDS / Salinity / DO Saturation) | 
pH
| Range | −2.000 to 20.000 pH | 
| Resolution | 0.1, 0.01, 0.001 pH | 
| Accuracy | ±0.002 pH | 
| Calibration Points | Up to 8 | 
| Standard Customization | Yes | 
| Calibration Reminder | Yes | 
| Standard Recognition | USA, NIST, MERK, JIS, GB, DIN buffers | 
| Slope Limit | Yes | 
mV / ORP
| Range | −2000.00 to 2000.00 mV | 
| Resolution | 0.1, 0.01 mV | 
| Accuracy | ±0.1 mV or ±0.03% | 
| EH ORP Mode | Yes | 
| ORP Calibration Points | 1 custom point (relative mV) | 
pX
| Range | −2.000 to 20.000 pX | 
| Resolution | 0.1, 0.01, 0.001 pX | 
| Accuracy | ±0.002 pX | 
| Calibration Points | Up to 6 | 
ISE
| Range | 1.000e−9 to 9.999e+9 | 
| Units | mol/L, mmol/L, g/L, mg/L, μg/L, ppm, ppb, pX | 
| Resolution | Up to 4 significant digits | 
| Accuracy | ±0.3% | 
| Calibration Points | Up to 8 | 
Conductivity
| Range | 0.000 μS/cm to 3000 mS/cm | 
| Resolution | 0.001 μS/cm minimum (varies with range selection) | 
| Accuracy | ±0.5% FS | 
| Reference Temperature | 5, 10, 15, 18, 20, 25 °C | 
| Calibration Points | Up to 5 | 
| Calibration Reminder | Yes | 
| Standard Recognition | 10 μS/cm, 84 μS/cm, 500 μS/cm, 1413 μS/cm, 12.88 mS/cm; 146.5 μS/cm, 1408 μS/cm, 12.85 mS/cm, 111.3 mS/cm | 
Resistivity
| Range | 5.00 Ω·cm ~ 100.0 MΩ·cm | 
| Resolution | 0.01 Ω·cm minimum | 
| Accuracy | ±0.5% FS | 
TDS
| Range | 0.000 ppm ~ 1000 ppt | 
| Resolution | 0.001 ppm minimum (changes with range) | 
| Accuracy | ±0.5% FS | 
Salinity
| Range | 0.0 ~ 80.0 ppt | 
| Resolution | 0.1 ppt | 
| Accuracy | ±1 ppt | 
Dissolved Oxygen — Concentration
| Sensor Type | Polarographic | 
| Range | 0.00 to 99.99 ppm | 
| Resolution | 0.01 ppm | 
| Accuracy | ±0.10 ppm (0.00–50.00), ±0.50 ppm (50.0–99.99) | 
| Calibration Reminder | Yes | 
| Calibration Points | Air-saturated water or zero point | 
| Barometric Compensation | Yes | 
| Manual Salinity Factor Correction | Yes | 
Dissolved Oxygen — Saturation
| Range | 0.0 to 600.0 % | 
| Resolution | 0.10 % | 
| Accuracy | ±2.0 % | 
Temperature
| Range | −10 to 135 °C, 14 to 275 °F | 
| Unit | °C, °F | 
| Resolution | 0.1 | 
| Relative Accuracy | ±0.1 | 
Measurement
| Reading Mode | Auto Read (Fast, Medium, Slow), Timed, Continuous | 
| Reading Prompts | Reading, Stable, Locked | 
| Temp. Compensation | ATC, MTC | 
Data management
| Data Storage | 1000 Groups | 
| GLP Features | Yes | 
Inputs & outputs
| pH Electrode | BNC (Q9) | 
| DO with Temp. Probe | 4-pin aviation connector | 
| Conductivity with Temp. Probe | 5-pin aviation connector | 
| Outputs | USB 2.0 flash memory device, PC, scanner | 
Display & power
| Backlight | Yes | 
| Auto Shutdown | 300, 600, 1200, 1800, 3600 s; Off | 
| Date & Time | Yes | 
| IP Rating | IP65 | 
General
| Power | Rechargeable Lithium battery; AC Adapter 100–240 V AC input, DC 5 V output | 
| Dimensions | 90 × 255 × 40 mm | 
| Weight | 500 g | 
 
There are no reviews yet.