Overview — VIKING-M510 portable multi-parameter meter
VIKING-M510 from COLO.Science is a rugged IP65 portable instrument for pH, mV/ORP, conductivity, resistivity, TDS, salinity, ISE, dissolved oxygen and temperature. It features a bright 4.3″ color touchscreen, an intelligent OS with user/calibration/electrode/method/data management, and GLP-compliant memory (1000 groups per parameter) for traceable results. Multi-reading modes (Auto-Read, Timed-Read, Continuous) and automatic/manual temperature compensation ensure reliable measurements in the lab or in the field.
Measurement modes & calibration
- Reading modes: Auto Read (Fast/Medium/Slow), Timed, Continuous
- pH: up to 8 points; USA/NIST/DIN/GB/MERK/JIS buffers; user-defined buffers; auto slope/offset diagnostics
- Conductivity: up to 5 points; ref. temp 5/10/15/18/20/25 °C; 10 μS/cm–12.88 mS/cm (and equivalents) standard recognition
- ISE: Direct, Standard Addition, Sample Addition, GRAN; >10 built-in methods (e.g., F−, Cl−, NO3−, NH4+, K+, Na+, Ca2+, Cu2+, Pb2+, Ag+); up to 8 points
- DO: polarographic sensor; air-saturated water or zero O2 calibration; auto barometric compensation; manual salinity correction
Key features
- 4.3″ high-contrast color touchscreen; backlight
- Intelligent OS: user / calibration / electrode / method / data / log management
- Auto-Hold endpoint detection; ATC/MTC
- USB 2.0 (flash drive / PC / printer-scanner)
- Reset to factory defaults; auto power-off (300–3600 s)
- IP65 enclosure for field and outdoor measurements
Inputs & outputs
- Inputs: pH electrode BNC (Q9); DO probe 4‑pin aviation; Conductivity probe 5‑pin aviation
- Outputs: USB 2.0 (flash drive, PC, scanner/printer)
Models & dimensions
Model |
Name |
Display |
IP Rating |
Dimensions (W×L×H) |
Weight |
VIKING-M510 |
Portable multi-parameter meter |
4.3″ color touchscreen |
IP65 |
90 × 255 × 40 mm |
≈ 500 g |
Technical specifications
Technical specifications — VIKING-M510 (Model M510T)
Model |
M510T |
Parameters |
pH / EC / ISE / DO / Temp. (mV / ORP / pX / Resistivity / TDS / Salinity / DO Saturation) |
pH
Range |
−2.000 to 20.000 pH |
Resolution |
0.1, 0.01, 0.001 pH |
Accuracy |
±0.002 pH |
Calibration Points |
Up to 8 |
Standard Customization |
Yes |
Calibration Reminder |
Yes |
Standard Recognition |
USA, NIST, MERK, JIS, GB, DIN buffers |
Slope Limit |
Yes |
mV / ORP
Range |
−2000.00 to 2000.00 mV |
Resolution |
0.1, 0.01 mV |
Accuracy |
±0.1 mV or ±0.03% |
EH ORP Mode |
Yes |
ORP Calibration Points |
1 custom point (relative mV) |
pX
Range |
−2.000 to 20.000 pX |
Resolution |
0.1, 0.01, 0.001 pX |
Accuracy |
±0.002 pX |
Calibration Points |
Up to 6 |
ISE
Range |
1.000e−9 to 9.999e+9 |
Units |
mol/L, mmol/L, g/L, mg/L, μg/L, ppm, ppb, pX |
Resolution |
Up to 4 significant digits |
Accuracy |
±0.3% |
Calibration Points |
Up to 8 |
Conductivity
Range |
0.000 μS/cm to 3000 mS/cm |
Resolution |
0.001 μS/cm minimum (varies with range selection) |
Accuracy |
±0.5% FS |
Reference Temperature |
5, 10, 15, 18, 20, 25 °C |
Calibration Points |
Up to 5 |
Calibration Reminder |
Yes |
Standard Recognition |
10 μS/cm, 84 μS/cm, 500 μS/cm, 1413 μS/cm, 12.88 mS/cm; 146.5 μS/cm, 1408 μS/cm, 12.85 mS/cm, 111.3 mS/cm |
Resistivity
Range |
5.00 Ω·cm ~ 100.0 MΩ·cm |
Resolution |
0.01 Ω·cm minimum |
Accuracy |
±0.5% FS |
TDS
Range |
0.000 ppm ~ 1000 ppt |
Resolution |
0.001 ppm minimum (changes with range) |
Accuracy |
±0.5% FS |
Salinity
Range |
0.0 ~ 80.0 ppt |
Resolution |
0.1 ppt |
Accuracy |
±1 ppt |
Dissolved Oxygen — Concentration
Sensor Type |
Polarographic |
Range |
0.00 to 99.99 ppm |
Resolution |
0.01 ppm |
Accuracy |
±0.10 ppm (0.00–50.00), ±0.50 ppm (50.0–99.99) |
Calibration Reminder |
Yes |
Calibration Points |
Air-saturated water or zero point |
Barometric Compensation |
Yes |
Manual Salinity Factor Correction |
Yes |
Dissolved Oxygen — Saturation
Range |
0.0 to 600.0 % |
Resolution |
0.10 % |
Accuracy |
±2.0 % |
Temperature
Range |
−10 to 135 °C, 14 to 275 °F |
Unit |
°C, °F |
Resolution |
0.1 |
Relative Accuracy |
±0.1 |
Measurement
Reading Mode |
Auto Read (Fast, Medium, Slow), Timed, Continuous |
Reading Prompts |
Reading, Stable, Locked |
Temp. Compensation |
ATC, MTC |
Data management
Data Storage |
1000 Groups |
GLP Features |
Yes |
Inputs & outputs
pH Electrode |
BNC (Q9) |
DO with Temp. Probe |
4-pin aviation connector |
Conductivity with Temp. Probe |
5-pin aviation connector |
Outputs |
USB 2.0 flash memory device, PC, scanner |
Display & power
Backlight |
Yes |
Auto Shutdown |
300, 600, 1200, 1800, 3600 s; Off |
Date & Time |
Yes |
IP Rating |
IP65 |
General
Power |
Rechargeable Lithium battery; AC Adapter 100–240 V AC input, DC 5 V output |
Dimensions |
90 × 255 × 40 mm |
Weight |
500 g |
There are no reviews yet.