Haze Measurement Guide — haze, transmittance, clarity and light transmittance

A clear film can let almost all light through and still look milky. A tinted window can look dark and still let heat in. Neither can be judged reliably by eye — they are measured. This haze measurement guide explains, in plain terms, what haze, total transmittance, clarity and light transmittance (VLT) mean, how a haze meter measures them, how to get repeatable results and how to choose an instrument.

It complements our Colour Measurement Guide, which covers colour of solids, liquids and powders. The two belong together: in plastics, glass and packaging, colour, gloss and haze describe the appearance of the same product.

1 · Haze, transmittance and clarity — what the numbers mean

When light hits a transparent or translucent sample, part is reflected at the surfaces, part is absorbed in the material and the rest passes through. Of the light that passes, some continues straight on and some is scattered by particles, crystallites, pigments, surface texture or scratches. The angle of that scattering decides what you see.

Haze measurement guide — direct, narrow-angle and wide-angle scattered light
Direct light, narrow-angle scattering (clarity) and wide-angle scattering (haze).
  • Total transmittance (T) — the share of incident light that passes the sample, direct and scattered together, in %. A clear 3 mm glass is typically around 90 %; the rest is lost mainly by reflection at the two surfaces.
  • Haze — the share of the transmitted light that is scattered by more than 2.5° from the incident beam, in %. High haze means a milky, veiled look and reduced contrast of objects behind the sample.
  • Clarity — related to scattering within a narrow angle, below 2.5°. Low clarity makes fine detail look blurred even when haze is low, typical for some films and textured surfaces.
  • Diffuse transmittance — the scattered part of the transmitted light; haze is diffuse transmittance divided by total transmittance.

Haze and transmittance are independent. A light-diffusing plate can have high transmittance and very high haze; a dark tinted glass can have low transmittance and almost no haze. That is why specifications usually state both.

2 · How a haze meter measures — 0/d geometry and the integrating sphere

A haze meter illuminates the sample with a parallel beam at 0° and collects the transmitted light in an integrating sphere — a hollow ball with a white, highly reflective inside. This is called 0/d geometry: 0° illumination, diffuse viewing. The sphere averages all transmitted light so that a single detector can measure it, regardless of the direction it was scattered in.

Haze meter 0/d geometry with integrating sphere, light trap and compensation port
0/d geometry: parallel beam, sample at the sphere entrance, exit port with light trap, detector.

Four readings, two results

Opposite the sample, the sphere has an exit port. When the port is closed with a white cap, all transmitted light is counted. When it is open into a light trap, the direct beam leaves the sphere and only the scattered light is counted. Combining readings with and without the sample gives:

ReadingSampleExit portWhat is measured
T1noclosed (white)Incident light — the 100 % reference
T2yesclosed (white)Total light transmitted by the sample
T3noopen (light trap)Light scattered by the instrument itself
T4yesopen (light trap)Light scattered by the sample and the instrument

Total transmittance T = T2 / T1. Diffuse transmittance Td = [T4 − T3 · (T2 / T1)] / T1. Haze = Td / T × 100 %. Modern instruments run this sequence automatically: after calibration you place the sample and read haze and T within seconds.

3 · ASTM D1003 or ISO 13468 / ISO 14782?

Both families of standards use 0/d geometry, but they treat the sphere differently. When the sample is placed at the entrance, it changes how much light the sphere wall returns — the sphere efficiency. The ASTM D1003 procedure accepts this; the ISO methods (ISO 13468 for total transmittance, ISO 14782 for haze) add a compensation port so that the sphere sees the same conditions with and without the sample.

ASTM D1003 (procedure A, haze meter)ISO 13468 / ISO 14782
Geometry0/d, integrating sphere0/d, integrating sphere
Sphere correctionNon-compensatedCompensation port
ResultsHaze, total transmittanceHaze, total luminous transmittance
Typical useNorth American and many global specificationsEuropean and ISO-based specifications
Results comparable?Close for low haze; differences can grow with high haze and thick or strongly scattering samples—

Use the method your customer or product specification names, and state it with the result (for example “haze 1.8 %, ASTM D1003, Ø20 mm aperture”). If you report to both, choose an instrument that supports both methods: CH1210, CH1610 and CH1810. CH1100 measures to ASTM D1003. The full texts of the standards are available from ASTM International and ISO.

4 · Clarity and the colour of transmitted light

Clarity

Haze tells you how milky a sample looks; clarity tells you how sharp an object looks when seen through it. Two films with the same haze can look very different when you read text through them. Clarity is evaluated from light scattered within a small angle around the direct beam and is reported in %. It is used for packaging films where the product must be clearly visible, for display and lens materials and for textured surfaces. Our CH1810 reports clarity together with haze and colour.

Colour of transmitted light

Transparent plastics yellow with processing temperature, UV exposure and ageing, and glass and films can carry a tint. A colour haze meter adds a spectral sensor (400–700 nm) and reports the colour of the transmitted light as CIE L*a*b*, yellowness index (YI), whiteness, Pt-Co or Gardner — in the same measurement as haze. See CH1610 and CH1810; for the theory of colour values see the Colour Measurement Guide.

5 · Measuring in practice — preparation, aperture, calibration

Sample preparation

  • Handle samples by the edges and keep them free of fingerprints, dust and protective film; fingerprints alone noticeably raise the haze value.
  • Clean with a soft, lint-free cloth only when the method allows it; solvents can attack plastic surfaces.
  • Condition plastic samples at the temperature and humidity the standard requires before measuring, typically 23 °C and 50 % relative humidity.
  • Samples must be flat and larger than the aperture, with parallel surfaces; mark the measuring position and the machine direction of films.

Choosing the aperture

Use the largest aperture that the sample allows: Ø20 mm averages more of the surface and gives the best repeatability; Ø8 or Ø4 mm are for small parts such as lenses and display windows. Compare results only at the same aperture.

Haze meter measuring apertures Ø20, Ø15, Ø8 and Ø4 mm

Vertical or horizontal

Sheets and films are measured vertically at the sphere port, held by a film clamp. Small parts, liquids in cells and powders are easier horizontally. An open measuring area allows large sheets to be measured without cutting.

Calibration and checks

  • Calibrate 100 % (open beam) and zero as the instrument requires, after warm-up and when ambient conditions change.
  • Check daily with a haze standard plate of a known value close to your samples; record the result to show the instrument is in control.
  • Measure each sample at several positions (for example 3–5) and report the mean; for films, rotate 90° to reveal orientation effects.
  • Report method, aperture, number of readings and sample thickness with every result.

6 · Common errors and how to avoid them

ErrorEffectRemedy
Fingerprints, dust, scratchesHaze too highHandle by edges, clean, inspect before measuring
Sample not flat or not at the portUnstable readings, haze too high or too lowUse the film clamp; press the sample flat against the port
Aperture larger than the sampleLight passes beside the sample — wrong T and hazeChoose a smaller aperture
Comparing different apertures or methodsApparent differences that are not realFix aperture and method in the specification
Single reading on an inhomogeneous filmPoor repeatabilityMeasure several positions and average
No calibration after warm-up or temperature changeDrift through the dayCalibrate and check with a standard plate
Textured or curved sample measured like a flat oneResults not comparableAgree a sample orientation and holder; compare like with like

7 · Applications — what is checked

ApplicationWhat is measuredWhy
Packaging films (PE, PP, PET, BOPP)Haze, T, clarityThe product must look clear and attractive through the film; haze also reveals processing problems
Optical and display films, cover glassHaze, T, colourContrast and brightness of the display; very low haze values
Anti-glare and matt surfacesHaze, THaze is the target property and must stay within a narrow window
Light diffusers, LED coversHaze, THigh haze for even light, high T for efficiency
Glass — clear, frosted, laminated, coatedT, haze, colourTransparency, privacy glass, ageing of laminates
Transparent moulded parts, lensesHaze, T, YIMoulding defects, yellowing, batch-to-batch control
Agricultural and greenhouse filmsT, hazeLight reaching the crop and how evenly it is spread
Liquids and resins in cellsHaze, T, colourTurbidity-like appearance and colour of clear liquids

Limit values depend on the product and are set in your specification; the instrument gives the comparable number.

8 · Light transmittance meters — VLT, UV and IR rejection, SHGC

Window film installers, glass processors and vehicle inspection need different numbers than a plastics laboratory: how much visible light the glazing lets through, and how much UV and heat it blocks. A compact light transmittance meter measures these in about half a second, directly on glass or film samples.

  • VLT — visible light transmittance: the share of visible light (380–760 nm) weighted to the sensitivity of the eye. Regulations for vehicle windows are based on it.
  • UV rejection: how much ultraviolet light is blocked — protection of interiors, goods and skin.
  • IR rejection: how much infrared is blocked at 940 nm, at 1400 nm or over the full IR band — the main lever for solar heat.
  • SHGC — solar heat gain coefficient: the share of solar energy that ends up as heat inside; lower is cooler.
  • TSER — total solar energy rejected: the share of solar energy kept out; higher is better.
Light transmittance meter for glass and window film

Vehicle glazing in Europe is regulated through UN Regulation No. 43, which sets a minimum visible light transmittance for the windscreen and front side windows; national rules for aftermarket film differ. VLT measured on the vehicle helps installers and inspectors verify compliance. CH1010 measures VLT, UV rejection and IR rejection at 940 nm; CH1020 adds IR at 1400 nm and full band, SHGC and TSER.

9 · Haze meter, transmittance meter or colour spectrophotometer?

Haze meter (0/d)Light transmittance meterBenchtop colour spectrophotometer
Main valuesHaze, total transmittance, clarity, colour of transmitted light (colour models)VLT, UV and IR rejection, SHGC, TSERColour (CIE Lab, ΔE) in reflectance and transmission; haze as an additional value
StandardsASTM D1003, ISO 13468, ISO 14782Instrument method for glazing and filmCIE colour standards; haze supplementary
SamplesFilms, sheets, glass, parts, liquids in cells; open measuring areaGlass, window film, glazing samplesSolids, powders, pastes, liquids
Best forLaboratory and production QC of transparent materialsInstallers, glass processing, inspectionColour QC where haze is a secondary value
Our instrumentsCH1100, CH1210, CH1610, CH1810CH1010, CH1020CS60 benchtop spectrophotometers

If haze is a value you sign certificates with, use a haze meter: the standard geometry, the large Ø154 mm sphere and the open sample area are designed for it.

10 · Choosing a model

CH1100CH1210CH1610CH1810
Haze, total transmittance✓✓✓✓
ISO method (compensation port)—✓✓✓
Colour of transmitted light——✓✓
Clarity———✓
Haze repeatability0.080.050.050.05
Bluetooth———✓
Sample thickness≤ 170 mm≤ 105 mm≤ 105 mm≤ 105 mm
  • Production control to ASTM D1003 — CH1100.
  • Laboratory reporting to ASTM and ISO — CH1210.
  • Haze plus yellowing or tint — CH1610.
  • Everything in one, including clarity — CH1810.
  • Glass and window film on site — CH1010; for solar-control film with SHGC and TSER — CH1020.

11 · Frequently asked questions

What is a good haze value?

It depends on the product. Optical films and display glass aim for very low haze, often below 1 %; packaging films typically a few percent; diffusers deliberately very high haze. The specification sets the limit.

Can I compare haze values measured on different instruments?

Yes, if method (ASTM or ISO), aperture and sample preparation are the same and both instruments are calibrated and checked with a standard plate. Our haze meters agree within 0.4 haze units on a standard plate at Ø20 mm.

Why does my haze change across a film roll?

Thickness, orientation, additives and surface condition vary across the web. Measure several positions, note the machine direction and compare like with like.

Does a haze meter measure turbidity of water?

Not in the sense of water-quality turbidity in NTU, which is measured with a turbidimeter. A haze meter can measure haze and transmittance of clear liquids in cells as a material property.

Is VLT the same as total transmittance?

Both describe light passing the sample. VLT is weighted to the eye over the visible range and is the value used for glazing; total transmittance from a haze meter is the standard value for plastics and films.