COLO.Science · Incubation & Controlled Environments · Shaker-Incubator
ORSHA75 · ORSHA125 Orbital Shaker-Incubator · Compact Benchtop · Exchangeable Platforms
The ORSHA75 orbital shaker-incubator and its larger sister ORSHA125 shake and incubate in one compact benchtop unit. A brushless DC motor drives the platform on a 20 mm orbit, and the chamber is held from 5 °C above room temperature up to 60 °C. ORSHA75 (PRO4550) has a 230 × 230 mm platform and runs at 50–300 rpm; ORSHA125 (PRO4559) has a 350 × 350 mm platform, 50–250 rpm and carries up to 10 kg.
ORSHA75 and ORSHA125 models and order codes
| Model | Order code | Platform size | Speed | Load | Exterior W × D × H | Power |
|---|---|---|---|---|---|---|
| ORSHA75 | PRO4550 | 230 × 230 mm | 50–300 rpm | 4.0 kg | 320 × 421 × 338 mm | 350 W |
| ORSHA125 | PRO4559 | 350 × 350 mm | 50–250 rpm | 10.0 kg | 632 × 502 × 512 mm | 600 W |
Both models heat from RT+5 °C to 60 °C with a 20 mm orbit and a brushless DC motor. The main unit is supplied without a platform: choose one or more platforms from the table below.
Compact orbital shaker-incubator for flasks, tubes and plates
Exchangeable platforms turn the same unit into a flask shaker, a tube incubator, a microplate incubator or a Petri dish incubator. The LCD shows the time, temperature and speed, and all three are set with a few keys. Heating and shaking stop as soon as the transparent lid is opened, and after a power cut the run continues by itself.

Key features
- Temperature control RT+5 °C to 60 °C, accuracy and uniformity ≤ ±0.5 °C at 37 °C
- Orbital shaking, 20 mm; ORSHA75 50–300 rpm, ORSHA125 50–250 rpm
- Direct brushless DC motor for gentle, reliable shaking
- LCD with time, temperature and speed; timer 1 min to 99 h 59 min or continuous
- Shaking and heating stop when the lid is opened
- Automatic restart after a power failure; over-heating protection
- Built-in temperature calibration
- Exchangeable platforms: 7 for ORSHA75, 4 for ORSHA125
Suitable for
- Shake-flask cultures of bacteria and yeast
- ELISA and deep-well plates incubated with shaking
- 15 ml and 50 ml tubes in clamp platforms
- Small laboratories and teaching, where one unit replaces a shaker and an incubator
Exchangeable platforms and order codes
The main unit is supplied without a platform. Choose the platform by the vessels you use; you can order several and change them in a minute. The spring holder is the most flexible choice for mixed flask and bottle sizes.
| Model | Platform | Order code | Holds |
|---|---|---|---|
| ORSHA75 | PF-1 | PRO4551 | 5 × 250 ml conical flasks (clamps) |
| ORSHA75 | PF-2 | PRO4552 | 9 × 100 ml conical flasks (clamps) |
| ORSHA75 | PF-3 | PRO4553 | 10 × 50 ml tubes (clamps) |
| ORSHA75 | PF-4 | PRO4554 | 18 × 15 ml tubes (clamps) |
| ORSHA75 | PF-5 | PRO4556 | spring holder, mixed vessels |
| ORSHA75 | PF-6 | PRO4557 | 4 ELISA or deep-well plates |
| ORSHA75 | PF-7 | PRO4558 | 5 Petri dishes (flat platform) |
| ORSHA125 | T1 | PRO4560 | 9 × 250 ml conical flasks (clamps) |
| ORSHA125 | T2 | PRO4561 | 25 × 100 ml conical flasks (clamps) |
| ORSHA125 | T3 | PRO4562 | 16 Petri dishes (flat platform) |
| ORSHA125 | T4 | PRO4563 | spring holder, mixed vessels |


Technical specifications
Temperature and time
| Temperature control | RT+5 to 60 °C |
| Setting range | 0 to 60 °C |
| Accuracy / uniformity | ≤ ±0.5 °C at 37 °C |
| Display resolution | 0.1 °C |
| Timer | 1 min to 99 h 59 min or continuous |
| Heating time to 60 °C | ≤ 15 min · ≤ 20 min |
Shaking and size (ORSHA75 · ORSHA125)
| Speed | 50–300 rpm · 50–250 rpm |
| Orbit | 20 mm |
| Platform | 230 × 230 · 350 × 350 mm |
| Chamber height | 182 mm · 325 mm |
| Load | 4 kg · 10 kg |
| Power | 350 W · 600 W |
| Weight | 18.5 kg · 41 kg |
| Supply | AC 220 V, 50/60 Hz |
Typical values. RT = room temperature. Exterior W × D × H: ORSHA75 320 × 421 × 338 mm, ORSHA125 632 × 502 × 512 mm.
Installation and care
Place the unit on a firm, level bench that does not vibrate, with free space around it for ventilation, and connect it to an earthed 220 V socket. Spread the vessels evenly over the platform and stay within the load limit. Wipe the chamber with a neutral detergent; do not use solvents on the acrylic lid.
Standard delivery and options
- ORSHA75 or ORSHA125 orbital shaker-incubator, main unit.
- Platform: ordered separately (see the platform table).
- Option: additional platforms for other vessels.
The configuration for a specific order is the one confirmed in the COLO.Science quotation.
Technical documents and knowledge base
ORSHA75 · ORSHA125 technical specification (PDF) — the orbital shaker-incubator in both sizes, with specifications, platforms and configuration checklist. Methods and application notes: our knowledge base.
How to choose
- ORSHA75 — compact, 320 × 421 mm of bench; up to 5 × 250 ml flasks, 4 microplates or 18 × 15 ml tubes.
- ORSHA125 — larger chamber, up to 9 × 250 ml or 25 × 100 ml flasks and 10 kg load.
- Need cooling below room temperature? See the ORSHAH / ORSHAHC orbital shaking incubators.
Configuration note
Tell us the vessels, the number per run, the temperature and the speed. COLO.Science will confirm the model and the platform in the quotation.
Sales and support for international customers
COLO Lab ExpertsPolje ob Sotli 4, SI-3255 Buče, Slovenia
Phone: +386 70 713 437 · Email: sales@colo.si · Web: colo.si
We supply laboratories across Europe directly and through partners. B2B prices are quoted net, excluding VAT.