Fluoride (F-) Combined Ion Selective Electrode


For fluoride in water · reference built in

F⁻ · 1 × 10⁻⁶ – 0.1 mol/L · Crystal membrane · 5–45°C · BNC (Q9)

PRO6237 — combined fluoride ion selective electrode: a combined fluoride ion selective electrode with a crystal membrane and a built-in refillable reference, so no separate reference electrode is needed. It measures fluoride (F⁻) from 1 × 10⁻⁶ – 0.1 mol/L and connects to the meter with a BNC (Q9) plug.

How the combined fluoride ion selective electrode is built

Polycarbonate
Body
Polycarbonate
Unbreakable shaft for daily handling. Not for organic solvents.
Ceramic
Junction
Ceramic
Good permeability, stable potential.
BNC (Q9)
Connector
BNC (Q9)
Fits ion meters and pH/mV meters with a BNC input.

The reference of the combined fluoride ion selective electrode is built in: an Ag/AgCl system behind a ceramic junction, refilled with 3 M KCl. No separate reference electrode is needed.

Technical parameters

Model / order code PRO6237
Electrode type Combined ion selective electrode
Measured ion F⁻
Measuring range 1 × 10⁻⁶ – 0.1 mol/L (19 µg/L – 1.9 g/L)
Membrane Crystal membrane
Body material Polycarbonate
Reference system Ag/AgCl
Junction Ceramic
Fill solution 3 M KCl
Interfering ions OH⁻
Working temperature 5 – 45 °C
Dimensions Φ12 × 120 mm
Connector BNC (Q9)

Full data sheet: Technical Specification (PDF), two pages.

Is this the right electrode for your sample?

The combined fluoride ion selective electrode measures fluoride ions in aqueous samples in the range 1 × 10⁻⁶ – 0.1 mol/L, which corresponds to 19 µg/L – 1.9 g/L, at 5 – 45 °C.

Interfering ions for this electrode are OH⁻; when they are present in the sample they affect the reading. Other ions and other designs are in the ion selective electrodes section of the electrode catalogue; the most frequent cases are:

  • Separate reference electrode already in the laboratory: the fluoride ion selective electrode as a half-cell.
  • Another ion: send us the ion and the sample type and we will check which ion selective electrode is available.
  • pH instead of ion concentration: a pH composite electrode from the same catalogue.

One electrode for benchtop and portable ion meters

The combined fluoride ion selective electrode does not belong to one meter. It works on a benchtop ion meter in the laboratory and on a portable one in the field — it needs a meter with an ion (concentration) or mV mode and a BNC input; see the pH/ion meters in the catalogue.

The response of an ion selective electrode depends on temperature, so standards and samples must be at the same temperature when they are measured. After moving the electrode to another meter, calibrate again: calibration data are stored in the meter, not in the electrode.

From sample to result

Sampling and keeping the sample

Take the sample in a clean plastic bottle, close it and measure it as soon as practical. Do not add acid or any other preservative to the portion used for the fluoride measurement, and let the sample reach the temperature of the standards before measuring.

Preparing the combined fluoride ion selective electrode before calibration and measurement

  • Remove the protective cap from the membrane at the bottom of the electrode and rinse the tip with distilled water; blot the drops, do not rub the membrane.
  • Check the level of the filling solution; if it is low, top up with 3 M KCl.
  • Leave the refill hole open during calibration and measurement.
  • Have the ionic strength adjuster ready: fluoride TISAB. It is added to every standard and every sample in the same proportion.

Calibrating the combined fluoride ion selective electrode

  1. Prepare at least two standards that bracket the expected concentration by diluting the 0.1 M fluoride standard (0.1 M NaF) step by step, for example tenfold.
  2. Add the ionic strength adjuster to each standard in the same proportion that will be used for the samples.
  3. Start the calibration on the meter and measure the standards from the lowest to the highest concentration; stir at the same gentle rate, wait for a stable reading and confirm each point.
  4. Rinse the tip with distilled water and blot the drops between solutions.
  5. The meter shows the electrode slope. If it is outside the limits given in the meter manual, clean the electrode, prepare fresh standards and repeat the calibration.

Measuring

Keep the refill hole open while measuring. Add the ionic strength adjuster to the sample in the same proportion as to the standards, stir at the same rate as during calibration, wait for a stable value and read the concentration. Rinse the tip with distilled water between samples.

Fluoride in drinking water

The combined fluoride ion selective electrode turns the meter into a direct-reading fluoride analyser: no colour reagent and no sample digestion, only the adjuster and a calibrated electrode.

Why it is measured: fluoride is followed in drinking water because both too little and too much of it matter for dental health, so waterworks and health laboratories check it routinely.

Why the ionic strength adjuster is needed: the fluoride buffer (TISAB) gives every standard and sample the same ionic strength, holds the pH where hydroxide — the interfering ion of this electrode — does not disturb the reading, and releases fluoride that is bound to aluminium and iron.

What the electrode reads: free fluoride ions in the solution; with the buffer added in the same proportion to standards and samples the reading corresponds to the fluoride concentration.

Ion meters, together with pH, conductivity and dissolved-oxygen meters, are in the electrochemistry catalogue.

Care, storage and accessories

Storing the electrode

After use, rinse the combined fluoride ion selective electrode with distilled water, blot it dry and put the protective cap back on the membrane, and close the refill hole. For storage between working days follow the instruction sheet supplied with the electrode.

When the reading is slow or drifts

The usual causes are a contaminated or scratched membrane, old or wrongly diluted standards, a missing ionic strength adjuster, a temperature difference between standards and samples, or a low level of filling solution. Clean the electrode, prepare fresh standards with the adjuster and calibrate again.

Accessories

Electrode holder
Electrode holder
Keeps the electrode at a fixed height in the sample and protects it.

See the electrode holder →

Electrode on the holder
Electrode on the holder
The arm lowers the electrode into the beaker and lifts it out for rinsing.

How the holder is used →

  • ISE standard: 0.1 M fluoride standard (0.1 M NaF), 250 ml.
  • Ionic strength adjuster: fluoride TISAB, 250 ml.
  • Filling solution: 3 M KCl for the built-in reference.

Video: how the electrode holder is installed and used

Standards, adjusters, filling solutions and holders are in the accessories section of the electrochemistry catalogue.

Frequently asked questions

Does the combined fluoride ion selective electrode fit my meter?

The combined fluoride ion selective electrode fits ion meters and pH/mV meters with a BNC input. Tell us the meter model before you order and we will confirm the connection.

How often should I calibrate the combined fluoride ion selective electrode?

Calibrate at the start of each working day and check the calibration with one standard every few hours; recalibrate when the check drifts or when the sample type or temperature changes.

Where can I read more about electrodes?

The electrode catalogue has an electrode guide, a full selection matrix and an overview of connectors. The electrochemistry overview covers pH, conductivity, dissolved oxygen and ion measurement.

Combined fluoride ion selective electrode and connector

Combined fluoride ion selective electrode PRO6237
PRO6237 · combined fluoride ion selective electrode · polycarbonate body, crystal membrane
BNC (Q9) plug
BNC (Q9) plug
Goes into the BNC input of the meter.
Not sure this is the right electrode for your sample?
Send us the meter model, sample type, expected concentration and sample pH — we will confirm the electrode or propose a better match from the catalogue.
Sales and support
COLO Lab Experts
Polje ob Sotli 4, SI-3255, Slovenia
Phone: +386 70 713 437 · Email: sales@colo.si · Web: colo.si
We supply laboratories and distributors across Europe. B2B prices are net, without VAT.
SKU: PRO6237 Category:

Request a Quote