COLO.SCIENCE – AUTOMATIC LABORATORY STEAM STERILIZATION
Internal water circulation and controlled liquid sterilization without fixed plumbing
The integrated SUS-304 water tank supplies steam to the chamber and recirculates water internally. The autoclave does not require fixed connections for water supply or drainage, supporting flexible installation in laboratories where a floor-standing sterilizer is needed.
Microprocessor control manages water filling, dynamic pulse exhaust, heating, sterilization, cooling and optional drying. The LCD displays pressure, temperature, time, cycle status, operating curves and fault information.
Key characteristics
- Models: VERSAC50, VERSAC75, VERSAC100 and VERSAC120.
- Nominal chamber volume: 50 / 75 / 100 / 120 L.
- Useful volume: 45 / 65 / 85 / 105 L.
- Sterilization temperature up to 133°C.
- Working pressure: 1.6 bar.
- LCD digital display and full microprocessor control.
- Ten selectable programs, including configurable cycles.
- Integrated SUS-304 water tank and inner circulation.
- USB and RS232 data connectivity.
Suitable for
- Liquid loads and thermolabile culture media.
- Agar preparation, dissolution and temperature holding.
- Packed and unpacked instruments.
- Dressings, equipment and rubber goods.
- Solid loads with rapid steam release.
- Microbiology, clinical, pharmaceutical and research laboratories.
- Waste sterilization and decontamination workflows.
VERSAC SERIES – VERTICAL STEAM STERILIZERS WITH INNER WATER CIRCULATION
VERSAC vertical autoclaves for liquids, culture media, instruments and laboratory waste
A VERSAC vertical autoclave is suitable when a laboratory needs automatic steam sterilization, an integrated water tank and flexible cycle programming without fixed water-supply and drainage connections. Four chamber sizes allow the sterilizer to be selected according to load volume, basket configuration, available installation height and electrical capacity.
Liquid programs can use rapid water cooling or variable natural cooling, while automatic selection of fast or slow exhaust helps reduce liquid overflow. An optional mobile probe can measure liquid temperature directly for application-specific cycle control.
For traceability, VERSAC autoclaves can be configured with a printer, USB and RS232 communication. Vacuum pulse, drying, Wi-Fi temperature display/alarm and accredited calibration are available according to the selected configuration and quotation.
VERSAC construction and sterilization technology

Floor-standing vertical design with a quick-opening hand-wheel door, self-expanding sealing ring and lockable caster wheels.
Liquid sterilization
Rapid water cooling and variable natural cooling support efficient liquid cycles. Automatic fast/slow exhaust selection helps protect loads against overflow.
Process monitoring
Dual temperature and pressure sensing supports controlled heating. The LCD shows process values, operating status, curves and diagnostic messages.
VERSAC series – model variants and selection parameters
VERSAC50
Nominal / useful volume: 50 / 45 L
Chamber: Ø410 × 450 mm
Baskets: 2 × Ø405 × 180 mm
Power: 2.5 kW
External size: 650 × 620 × 800 mm
Net weight: 87 kg
VERSAC75
Nominal / useful volume: 75 / 65 L
Chamber: Ø410 × 600 mm
Baskets: 2 × Ø405 × 270 mm
Power: 2.7 kW
External size: 650 × 620 × 900 mm
Net weight: 95 kg
VERSAC100
Nominal / useful volume: 100 / 85 L
Chamber: Ø410 × 650 mm
Baskets: 2 × Ø405 × 295 mm
Power: 3.0 kW
External size: 650 × 620 × 1000 mm
Net weight: 136 kg
VERSAC120
Nominal / useful volume: 120 / 105 L
Chamber: Ø410 × 1055 mm
Baskets: 3 × Ø405 × 310 mm
Power: 4.5 kW
External size: 650 × 620 × 1370 mm
Net weight: 151 kg
Model note: final chamber configuration, basket quantity, electrical supply, accessories and documentation must be confirmed in the official quotation.
TECHNICAL SPECIFICATIONS – VERSAC VERTICAL AUTOCLAVE SERIES
| Product series | VERSAC vertical steam sterilizers |
| Model variants | VERSAC50 / VERSAC75 / VERSAC100 / VERSAC120 |
| Chamber volumes | 50 / 75 / 100 / 120 L |
| Useful volumes | 45 / 65 / 85 / 105 L |
| Chamber material | Medical-grade SUS-304 stainless steel |
| Maximum working pressure | 1.6 bar |
| Working temperature range | 105-128°C |
| Maximum stated sterilization temperature | Up to 133°C, program/configuration dependent |
| Temperature display accuracy | ±0.1°C |
| Temperature variation | ≤ ±0.1°C |
| Temperature sensor | Pt100, Class A |
| Sterilization timer | 1-250 min |
| Variable timer | 1 min to 7 days |
| Dynamic pulse exhaust | 0-9 cycles |
| Display and control | LCD digital display, full microprocessor control |
| Water requirement | Distilled or demineralized water |
| Melting range | 60-100°C, programmable |
| Heating/holding range | 45-60°C, programmable |
| Electrical supply | AC 230/240 V, 50/60 Hz, single phase; 14.5 A fuse |
| Communication | USB and RS232 |
| Installation | Integrated water tank; no fixed water-supply or drainage connection required |
The exact operating range, process options, cooling method, accessories and final delivered configuration are defined in the official quotation and order documentation.
STANDARD DELIVERY / OPTIONAL CONFIGURATION
VERSAC standard equipment and optional accessories
Standard delivery
- Selected VERSAC autoclave.
- Model-specific stainless-steel baskets.
- Removable power connection.
- User documentation.
- Integrated stainless-steel water tank.
Optional accessories and services
- Built-in thermal printer.
- Mobile liquid-temperature probe.
- Vacuum-pulse sterilization.
- Drying function.
- Wi-Fi temperature display/alarm.
- Accredited pressure, temperature and safety-valve calibration.
Sterilization programs and process workflow
Solid and porous loads
Programs for packed and unpacked instruments, dressings, equipment, rubber goods and solids with rapid steam release.
Liquids and culture media
Liquid, custom liquid, culture-media, agar preparation, media dissolution and temperature-holding programs.
Test and waste programs
Bowie-Dick testing and laboratory-waste sterilization are included among the selectable cycle applications.
Data and scheduling
Automatic start, timer programming, USB/RS232 export and an optional printer support process records and scheduled laboratory operation.
MULTI-LAYER PROCESS AND OPERATOR PROTECTION
VERSAC safety systems
Pressure and temperature protection
- Overpressure alarm and automatic safety-valve release.
- Overtemperature alarm with automatic power-off.
- Independent safety thermostat and circuit breaker.
- Electrical and mechanical pressure indication.
Door and water protection
- Door/lid status sensor and electromagnetic safety lock.
- Hydraulic locking while pressure remains in the chamber.
- Anti-dry-out protection and water-level monitoring.
- Thermally insulated safety cover over the lid.
Process diagnostics
- Chamber and water-tank sensors.
- Visual and audible malfunction alarms.
- Error-code display on the LCD.
- Open-lid sensing and controlled cycle interruption.
AUTOCLAVE CATALOGUE
Compare vertical, benchtop and large-capacity autoclaves
Use the COLO product catalogue and autoclave category to compare chamber volume, control systems, liquid-cycle options, cooling, drying, traceability and accessories.
COLO.SCIENCE KNOWLEDGE AND SUPPORT
Selection support for steam sterilization, liquid cycles and validation
COLO.Science supports model selection, load definition, cycle configuration, cooling method, printer/data workflow, calibration and final documentation requirements.
How to select the appropriate VERSAC model
First define the usable chamber volume and the physical load: basket quantity, flask or container height, number of vessels, total load mass and required free steam-circulation space. VERSAC50 and VERSAC75 suit smaller routine batches, while VERSAC100 and VERSAC120 provide more capacity for larger laboratories and higher-throughput workflows.
Next define the load type and process: solid, porous, liquid, culture media, agar, waste or a mixed laboratory workflow. Liquid loads may require a mobile probe, controlled exhaust, rapid water cooling or natural cooling according to container type and process requirements.
Finally confirm electrical supply, printer and data export, drying or vacuum-pulse options, calibration scope, regulatory documentation and installation access. The final quotation should state the exact model and delivered configuration.
VERSAC vertical steam sterilizers for automatic laboratory autoclaving
VERSAC autoclaves combine an integrated stainless-steel water tank, automatic process control and model capacities from 50 to 120 L. Their flexible programs support instrument, dressing, rubber, liquid, culture-media, agar and waste applications in biomedical, pharmaceutical, food, clinical and research laboratories.
In practical use, cycle selection depends on load composition, container type, fill volume, heat sensitivity, required cooling method, drying requirement and documentation workflow. Printer, mobile probe, Wi-Fi alarm, vacuum pulse, drying and accredited calibration can be selected according to the application.
Configuration notice
VERSAC vertical autoclaves may be supplied in different configurations depending on selected model, sterilization load, cooling method, liquid probe, printer, USB/RS232 workflow, Wi-Fi alarm, vacuum pulse, drying, calibration and documentation requirements. The official quotation defines the final delivered configuration. Specifications may change without notice.