Two Shops
One Shipment

CIBAC45 Cooling Circulating Bath −10 to 100 °C, 7.3 L, Touch Screen


COLO.Science · Cooling Circulating Bath

−10–100 °C · ±0.1 °C · 7.3 L · 0–15 L/min · Touch screen

COLO.Science · Heating & Thermal Processing · Circulating Bath

CIBAC45 Cooling Circulating Bath · −10 to 100 °C · 7.3 L · ±0.1 °C

The cooling circulating bath CIBAC45 cools and heats liquid from −10 °C to 100 °C and holds it within ±0.1 °C. The 7.3 L stainless steel tank takes samples directly, and the pump sends the liquid through hoses to a rotary evaporator condenser, a jacketed reactor or another instrument that needs a constant temperature.

Cooling circulating bath for the bench

The CIBAC45 cooling circulating bath combines a compressor, a heater and a circulation pump in one unit. Below room temperature it replaces tap water or ice for cooling a condenser; above it, the same bath works as a heated water or oil bath up to 100 °C. Switch between internal circulation, which keeps the tank uniform, and external circulation, which carries the liquid to a connected vessel and back.

CIBAC45 cooling circulating bath 7-inch touch screen controller

Key features

  • Temperature −10 °C to 100 °C, stability ±0.1 °C, resolution 0.1 °C
  • Stainless steel tank 260 × 200 × 140 mm, 7.3 L, with lid and drain valve
  • Pump 0–15 L/min, up to 0.45 bar, internal or external circulation
  • 7-inch touch screen with set and actual temperature, time and status
  • Programs up to 30 segments, timer, settings kept after a power failure
  • Alarms for over- and under-temperature and low liquid level; compressor overload protection

Suitable for

  • Cooling the condenser of a rotary evaporator
  • Heating and cooling jacketed reactors and vessels
  • Cooling samples during ultrasonic processing
  • Constant temperature for samples, reagents and test pieces in the tank

CIBAC series: four tank sizes, down to −40 °C

Tank Tank size (mm) Models and lowest temperature (upper limit 100 °C)
7.3 L 260 × 200 × 140 CIBAC45 −10 °C · CIBAC46 −20 °C · CIBAC47 −30 °C · CIBAC48 −40 °C
10 L 260 × 200 × 200 CIBAC49 −5 °C · CIBAC50 −10 °C · CIBAC51 −20 °C · CIBAC52 −30 °C · CIBAC53 −40 °C
15 L 300 × 250 × 200 CIBAC54 −5 °C · CIBAC55 −10 °C · CIBAC56 −20 °C · CIBAC57 −30 °C · CIBAC58 −40 °C
29 L 440 × 325 × 200 CIBAC59 −5 °C · CIBAC60 −10 °C · CIBAC61 −20 °C · CIBAC62 −30 °C

All cooling circulating bath models: stability ±0.1 °C, pump 0–15 L/min, 7-inch touch screen. Order codes, power and outer dimensions of every model are in the CIBAC technical specification (PDF).

Which model to choose

To choose a cooling circulating bath, start with the lowest temperature you need, then the tank size. CIBAC45 covers −10 °C with a 7.3 L tank; with the same tank CIBAC46 goes to −20 °C, CIBAC47 to −30 °C and CIBAC48 to −40 °C. For more liquid or larger vessels choose the 10 L, 15 L or 29 L models. When you only need to heat, see the PURUS thermostatic water bath, and when the samples must also be shaken the ORSHAWB shaking water bath.

Technical specifications

Temperature and control

Temperature −10 to 100 °C
Stability ±0.1 °C
Resolution 0.1 °C
Control PID, program up to 30 segments
Display 7-inch touch screen
Pump 0–15 L/min, max. 0.45 bar

Tank and construction

Tank 7.3 L, stainless steel
Tank size 260 × 200 × 140 mm
Top opening 180 × 140 mm
Heating power 1.0 kW
Total power 1.3 kW
Outer size 390 × 340 × 652 mm
Supply 220 V, 50 Hz

Values measured without external load, at an ambient temperature of 25 °C. Ask us for cooling capacity, packing size and weight.

Installation and bath liquid

Leave 300 mm free space behind and on both sides of the cooling circulating bath for ventilation and connect it to an earthed socket. Keep the liquid at least 30 mm below the top of the tank. Use ethanol or a water–glycol mixture below 5 °C, pure water from 5 to 80 °C, water with about 15 % glycerol from 80 to 90 °C and silicone oil above 90 °C.

Standard delivery

  • CIBAC45 cooling circulating bath with stainless steel tank, 7.3 L.
  • Stainless steel lid.
  • Power cord and rubber hose.
  • Operating manual.

The configuration for a specific order is the one confirmed in the COLO.Science quotation.

Technical documents and knowledge base

CIBAC technical specification (PDF) — CIBAC45 data, the full CIBAC series table with order codes, bath liquids and configuration checklist. Methods and application notes: our knowledge base.

Configuration note

To configure a cooling circulating bath, tell us the lowest and highest working temperature and what you cool or heat — samples in the tank or an external condenser or reactor. COLO.Science will confirm the model and the delivery time in the quotation.

Sales and support for international customers

COLO Lab ExpertsPolje ob Sotli 4, SI-3255 Buče, Slovenia
Phone: +386 70 713 437 · Email: sales@colo.si · Web: colo.si
We supply laboratories across Europe directly and through partners. B2B prices are quoted net, excluding VAT.

Need to cool a condenser or a reactor?

Send the working temperature range and what you cool or heat. COLO.Science will confirm the model and the delivery time.

Thermostatic Baths

, ,