Technical Specification
COLO.Science refrigerated centrifuge

LACE56R

High Capacity Refrigerated Benchtop Centrifuge | -20 °C to +40 °C

LACE56R combines the flexible LACE56 rotor platform with active refrigeration for temperature-sensitive laboratory samples. It delivers up to 6000 rpm, 6321 x g and 4 × 750 ml while maintaining a selectable chamber temperature from -20 °C to +40 °C with ±1 °C precision.

Fast refrigeration: cools to 4 °C in 8 minutes, reducing preparation time between runs and helping protect enzymes, proteins, blood components, vaccines and other temperature-sensitive materials.
6000 rpm6321 x g4 × 750 ml max.-20 °C to +40 °C4 °C in 8 min5-inch touch screenAuto rotor ID
Document: English Technical Specification
Product: LACE56R refrigerated benchtop centrifuge
Rotor families: same LACE56 platform portfolio | Temperature precision: ±1 °C
Format: A4 landscape | 7 pages | print-ready HTML
LACE56R refrigerated benchtop centrifuge
LACE56R refrigerated high-capacity centrifuge | rotor configured according to order
Product No.PRO6992
ModelLACE56R
Cooling-20 to +40 °C
Page 1 / 7
COLOLACE56R | Technical Overview
Refrigerated model ยท PRO6992
Rotor-dependent performance
Temperature protection, throughput and controlHigh-capacity refrigerated centrifugation with wide rotor and adapter selection
Maximum speed6000 rpm
Maximum RCF6321 x g
Max. capacity4 × 750 ml
Temperature-20 to +40 °C
Cooling4 °C in 8 min
Refrigeration and sample protection
Rapid pre-cooling: reaches 4 °C in 8 minutes to shorten setup and turnaround between temperature-controlled runs.
±1 °C temperature precision: limits chamber temperature variation during protocols involving proteins, enzymes, blood fractions and other sensitive materials.
Wide temperature range: -20 °C to +40 °C supports chilled separation, controlled ambient operation and protocol flexibility.
Large-batch protection: up to 4 × 750 ml allows high-throughput processing without giving up active cooling.
Imported fast-cooling system: designed for stable, repeated refrigerated laboratory operation.
Important: chamber set temperature is not automatically the same as actual sample temperature. Validate critical protocols using the vessel, load, rotor, run time and starting temperature used in practice.
Control, safety and workflow
Automatic rotor identification: supports correct rotor limits and helps reduce setup errors.
5-inch touch panel: clear access to speed, RCF, time, temperature and ramp parameters, including adjustment during operation.
10 acceleration / deceleration profiles: gentler ramps help protect fragile pellets, gradients and interfaces from disturbance.
Multiple working modes: timed, differential and gradient centrifugation support routine and multi-step workflows.
Protection system: electronic lid lock, imbalance stop, over-speed and over-temperature protection, motor over-current protection and error diagnostics.
NCPNCP Rotor TechnologyCompatible rotor options remain shared with the LACE56 platform.
Configuration note: select the rotor by vessel type, volume, bottom geometry, target RCF, cooling requirement and biosafety sealing need. The lowest-rated rotor, bucket, adapter or vessel determines the permitted operating limit.
Page 2 / 7
COLOLACE56R | Main Technical Data
Refrigerated benchtop model
PRO6992 host centrifuge
Main technical specifications
ParameterSpecification
ModelLACE56R refrigerated benchtop centrifuge
Model code referencePRO6992
Maximum speed6000 rpm, rotor dependent
Maximum RCF6321 x g, rotor dependent
Maximum capacity4 × 750 ml with compatible swing-out rotor
Temperature range-20 °C to +40 °C
Temperature precision±1.0 °C
Cooling performanceReaches 4 °C in 8 minutes
Control interface5-inch touch panel
Rotor identificationAutomatic
ParameterSpecification
Power supplyAC 220 ±22 V, 50 Hz, 10 A
Total power1200 W
Speed accuracy±10 rpm
Timer range1 s - 99 min 59 s / 1 min - 99 h 99 min
Acceleration / deceleration10 profiles
Working modesTimed / differential / gradient centrifugation
Noise level≤65 dB(A)
Net weight85 kg
Dimensions (W × D × H)640 × 610 × 340 mm
Rotor / adapter range0.2 ml to 750 ml vessels, microplates and selected blood tube / blood bag workflows
Configuration package logic
Host centrifuge
PRO6992 LACE56R base unit
Active refrigeration, -20 °C to +40 °C
Rotor configured separately
Rotor family
Same LACE56 platform rotor portfolio
4 × 250, 300, 500 and 750 ml swing-out rotors
Microplate and blood tube rotors
Application fit
Clinical and blood testing
Biopharmaceutical and cell culture workflows
RIA, molecular chemistry and biosafety laboratories
Specification note: maximum speed, RCF and capacity are not necessarily achieved by the same rotor. Confirm the selected rotor, adapters and vessel ratings in the binding quotation and rotor documentation.
Page 3 / 7
COLOLACE56R | Rotor Portfolio 1
Rotor cards with compatible adapter chips
NCP badges added where applicable
Rotor portfolioPart 1 of 3 | compatible adapters are shown directly on each rotor card
12x10 ml Rotor (Rounded) 6000rpm/6321xg
SKU PRO6945
NCP

12x10 ml Rotor (Rounded) 6000rpm/6321xg

Listed limit: 6000 rpm / 6321 x g
Compatible adapters
Direct rotor configuration
4x250ml Swing out rotor (Square cup) 4000rpm/3040xg
SKU PRO6946

4x250ml Swing out rotor (Square cup) 4000rpm/3040xg

Listed limit: 4000 rpm / 3040 x g
Compatible adapters
PRO6947 4x4x50ml (Conical bottom)PRO6948 4x8x15ml (Conical bottom)PRO6949 4x8x10ml (rounded bottom)PRO6950 4x14x5ml (rounded bottom)PRO6951 4x14x1.5/2ml (rounded bottom)PRO6952 4x10x10ml (blood collection tube)PRO6953 4x18x5ml (blood collection tube)PRO6959 4x4x50ml (Conical Bottom)
4×300ml Swing out rotor (4x250ml Conical -bottom Corning bottle) 4000rpm/3040xg
SKU PRO6954

4×300ml Swing out rotor (4x250ml Conical -bottom Corning bottle) 4000rpm/3040xg

Listed limit: 4000 rpm / 3040 x g
Compatible adapters
PRO6955 50 to 300ml for rotor PRO6954 volume at choice
4×500ml Swing out Biosafety Rotor (Rounded) 4000rpm/3500xg
SKU PRO6956
BioSafe

4×500ml Swing out Biosafety Rotor (Rounded) 4000rpm/3500xg

Listed limit: 4000 rpm / 3500 x g
Compatible adapters
PRO6958 4x250ml (Rounded)PRO6959 4x4x50ml (Conical Bottom)PRO6960 4x100ml (Rounded)PRO6961 4x9x15ml (Conical Bottom)PRO6962 4x9x10ml (Rounded)PRO6963 4x14x5ml (Rounded)PRO6964 4x14x1.5/2mlPRO6965 4×12×10ml (Blood collection tube Cap-removing)PRO6966 4×14×2/5/7 ml (Blood collection tube Cap-removing)
Page 4 / 7
COLOLACE56R | Rotor Portfolio 2
High-capacity swing-out rotor options
Adapter chips visible on each card
Rotor portfolioPart 2 of 3 | large bottle and tube formats
4×500ml Swing out Rotor (Rounded) 4000rpm/3500xg
SKU PRO6957

4×500ml Swing out Rotor (Rounded) 4000rpm/3500xg

Listed limit: 4000 rpm / 3500 x g
Compatible adapters
PRO6958 4x250ml (Rounded)PRO6959 4x4x50ml (Conical Bottom)PRO6960 4x100ml (Rounded)PRO6961 4x9x15ml (Conical Bottom)PRO6962 4x9x10ml (Rounded)PRO6963 4x14x5ml (Rounded)PRO6964 4x14x1.5/2mlPRO6965 4×12×10ml (Blood collection tube Cap-removing)PRO6966 4×14×2/5/7 ml (Blood collection tube Cap-removing)
4×750ml Swing out Rotor (Rounded) 4000rpm/3500xg
SKU PRO6967

4×750ml Swing out Rotor (Rounded) 4000rpm/3500xg

Listed limit: 4000 rpm / 3500 x g
Compatible adapters
PRO6968 4x500ml (Conical /Round Bottom)PRO6969 4x250 ml (Conical /Round Bottom)PRO6970 4x3x100 ml (Conical /Round Bottom)PRO6971 4x7x50 ml (Conical)PRO6972 4x14x15 ml (Conical)PRO6973 4x14x10 ml (Rounded)PRO6974 4x21x5/6 ml (Rounded)PRO6975 4x26x1.5/2 ml (Conical/Rounded)PRO6976 4x19x10 ml (Blood collection tube Cap removing)PRO6977 4x24x5/7 ml (Blood collection tube Cap removing)
4×750ml Swing out Rotor (Square) 4000rpm/3580xg
SKU PRO6978

4×750ml Swing out Rotor (Square) 4000rpm/3580xg

Listed limit: 4000 rpm / 3580 x g
Compatible adapters
PRO6979 4×500ml (Rounded)PRO6980 4×250ml (Rounded)PRO6981 4×3x100ml (Rounded)PRO6982 4×7x50ml (Conical)PRO6983 4×14x15ml (Conical)PRO6984 4×14x10ml (Rounded)PRO6985 4×24x5/6ml (Rounded)PRO6986 4×30x1.5/2ml (Rounded/Conical)PRO6987 4×24x10ml (Blood collection tube cap removing)PRO6988 4×30x2/5/7ml (Blood collection tube cap removing)
Swing out Rotor Microplate Holder Rack 4x4x96 4000rpm/3040g
SKU PRO6989

Swing out Rotor Microplate Holder Rack 4x4x96 4000rpm/3040g

Listed limit: 4000 rpm / 3040 x g
Compatible adapters
Direct rotor configuration
Page 5 / 7
COLOLACE56R | Rotor Portfolio 3
Microplate and blood tube rotor options
Final quotation must confirm included parts
Rotor portfolioPart 3 of 3 | microplate, blood tube and remaining formats
96x5/7 ml Swing out Rotor Blood collection Tube 4000rpm/3000xg
SKU PRO6990

96x5/7 ml Swing out Rotor Blood collection Tube 4000rpm/3000xg

Listed limit: 4000 rpm / 3000 x g
Compatible adapters
Direct rotor configuration
120x5/7 ml Swing out Rotor Blood collection Tube 4000rpm/3500xg
SKU PRO6991

120x5/7 ml Swing out Rotor Blood collection Tube 4000rpm/3500xg

Listed limit: 4000 rpm / 3500 x g
Compatible adapters
Direct rotor configuration
12×10 ml Angle Rotor 6000 rpm / 6321 xg
SKU PRO7063
NCP

12×10 ml Angle Rotor 6000 rpm / 6321 xg

Listed limit: 6000 rpm / 6321 x g
Compatible adapters
Direct rotor configuration
Configuration note: special-format rotors should be selected according to vessel geometry, loading pattern and required operating speed. Confirm final product photography before catalogue or quotation use.
Page 6 / 7
COLO.ScienceLACE56R Product Support
COLO.Science | Centrifugation support
Manufacturer and support
COLO Lab Experts
Polje ob Sotli 4, SI-3255, Slovenia
LACE56R selection should be matched to sample volume, bottle format, tube bottom geometry, target RCF, balancing requirements, biosafety expectation and the required rotor/adaptor combination.
Binding specification notice: This document supports product selection and does not replace the binding quotation, order confirmation, manufacturer manual or rotor certificate. Rotor performance is configuration dependent. Confirm SKU, included rotor, buckets, adapters, vessel dimensions, maximum permitted speed/RCF, safety documentation and final images before ordering or operation. Specifications may change without notice.
Page 7 / 7