Precise refrigeration control
The temperature setting range extends from −20 to +40 °C with accuracy of ±1.0 °C, supporting temperature-sensitive separation across high- and low-speed workflows.
Refrigerated high- and low-speed centrifuge · SKU PRO7073
Versatile refrigerated high- and low-speed laboratory centrifuge engineered for sample volumes from 2 to 250 ml, with precise temperature and speed control, rapid cooling, intelligent method storage and a corrosion-resistant nano-ceramic-coated chamber.
The temperature setting range extends from −20 to +40 °C with accuracy of ±1.0 °C, supporting temperature-sensitive separation across high- and low-speed workflows.
Under the stated test conditions, the chamber cools from 25 °C to 4 °C in approximately six minutes using an environmentally considerate refrigerant.
HD touch-panel control with storage for 99 custom programs and USB data export supports method recall and centrifugation-record documentation.
Interchangeable angle and swing-out rotors support high-speed microtube processing and lower-speed, higher-capacity workflows with vessels up to 250 ml.
Ten acceleration/deceleration levels, soft start/stop control, three-stage rubber vibration absorption and automatic balancing protect samples and support noise levels up to 60 dB(A).
Selected carbon nano-polymer composite rotors combine low mass, high strength, low thermal conductivity and corrosion resistance. The nano-ceramic chamber coating adds protection against corrosion and rust.
Rotor, tube and speed must always be selected according to the validated method, sample characteristics and tube manufacturer's limits.
Configuration notice: RITACUBA20R is supplied as a refrigerated centrifuge unit. The rotor and any required adapter are selected separately according to tube format, sample volume, target RCF, temperature requirement and workflow. Maximum instrument performance depends on the installed rotor; never exceed the limits stated for the rotor, adapter or tube.
6,000 rpm · 6,321 × g
10 ml rounded-bottom tubes
NCP COMPOSITE ROTOR
16,500 rpm · 21,610 × g
Microtubes: 1.5/2.0 ml
NCP COMPOSITE ROTOR
15,000 rpm · 22,387 × g
Microtubes: 1.5/2.0 ml
NCP COMPOSITE ROTOR
15,000 rpm · 22,262 × g
Microtubes: 1.5/2.0 ml
NCP COMPOSITE ROTOR
12,000 rpm · 16,260 × g
Microtubes: 1.5/2.0 ml
NCP COMPOSITE ROTOR
15,000 rpm · 24,148 × g
5 ml round-bottom tubes
NCP COMPOSITE ROTOR
11,000 rpm · 15,620 × g
15 ml conical-bottom tubes; optional 12 × 5 ml adapter
NCP COMPOSITE ROTOR
11,000 rpm · 14,440 × g
50 ml Falcon-type tubes; configurations for 6 × 15 ml conical or 6 × 10 ml round-bottom tubes
NCP COMPOSITE ROTOR
4,000 rpm · 3,040 × g
Square cups; adapters support 50, 15, 10, 5 and 1.5 ml tubes and blood collection tubes
11,000 rpm · 15,354 × g
Adapters for 6 × 50 ml, 6 × 15 ml and 6 × 3 × 10 ml tubes
NCP COMPOSITE ROTOR4,000 rpm · 2,683 × g
Available adapters: 4 × 50 ml and 8 × 15 ml + 1.5 ml
Provides controlled access to samples after a power interruption, following the operating manual and only after the rotor has fully stopped.
Timed, inching, differential-time and gradient centrifugation modes support routine separation and more specialised refrigerated protocols.
A one-turn NCP rotor-lid design, aerodynamic buckets, three-stage rubber shock absorption and automatic balancing support convenient, stable operation.