COLO.Science · Technical Specification
VIKING-ISE500
Benchtop pH / Ion Meter · ISE / ORP / mV / Temperature · 7.0″ Touchscreen
VIKING-ISE500 Benchtop pH/Ion Meter is a professional modular electrochemistry analyzer for high-accuracy measurement of pH, ion-selective electrodes (ISE), oxidation-reduction potential (ORP), mV and temperature. The instrument combines a 7.0″ color touchscreen, intelligent method and user management, GLP data functions, electrode diagnostics, USB / RS-232 connectivity and IP54 protection for demanding laboratory and water-quality workflows.
Benchtop pH / ion meter pH / ISE / ORP / mV Temperature 7.0″ touchscreen ±0.002 pH 1–6 point ISE calibration GLP data management USB / RS-232 1000 data sets / parameter IP54
Document type: Technical Specification Sheet
Product family: Benchtop Electrochemistry / Water Quality Meter
Model: VIKING-ISE500 · SKU: PRO2561
Expected HTML path: colo.si/wp-content/images/TechSpec/WaterQualityP/VIKING-ISE500.html
Expected PDF path: colo.si/wp-content/images/TechSpec/WaterQualityP/VIKING-ISE500.pdf
Format: A4 landscape · COLO.Science TechSpec v2.4 with navigation links
VIKING-ISE500 benchtop pH ion meter main unit SKU PRO2561
VIKING-ISE500 · benchtop pH / ISE / ORP / mV / temperature meter · main unit
Model / SKU VIKING-ISE500
PRO2561
Address / HQ Polje ob Sotli 4
SI-3255, Slovenia
Page 1 / 5
COLO.Science logo VIKING-ISE500 · Benchtop pH / Ion Meter
Technical data sheet · Informational document · Controlled electronic document
Product category: Water quality instruments on colo.si
VIKING-ISE500 · technical specification Benchtop analyzer for pH, ion-selective electrodes, ORP, mV and temperature measurement
pH -2.000 to 20.000 ISE 1.00E-9 to 9.99E9 ORP / mV ±2000.00 Temp. -5 to 130°C Direct / addition / Gran plot — Modular benchtop platform for high-precision pH and ion measurement
📋 Product Overview
Application profile
Ion-selective electrode work: supports direct reading, standard addition, sample addition and Gran plot techniques for ISE workflows.
Water quality laboratories: suitable for fluoride, ammonium, nitrate and other ion methods where appropriate electrodes and standards are selected.
Research and method development: flexible method, electrode and user management for laboratories developing or validating electrochemical procedures.
Routine QC: combines pH, ORP / mV and temperature measurement with GLP data storage and stable reading modes.
Design and handling advantage
Touchscreen workflow: 7.0″ color touchscreen supports intuitive operation, method selection and measurement review.
Traceability: user management, calibration management, method management, data review and audit-oriented functions support controlled laboratory work.
Electrode diagnostics: automatic slope and offset display, calibration reminder and slope-limit alert help maintain measurement reliability.
Connectivity: USB and RS-232 interfaces support data transfer, printing or connection to laboratory reporting workflows.
Key features
pH performance: -2.000 to 20.000 pH range, 0.001 pH maximum resolution and ±0.002 pH accuracy.
ISE performance: 1.00E-9 to 9.99E9 range with 4 significant digits and ±0.3% accuracy.
ISE calibration: 1–6 calibration points with support for direct reading, standard / sample addition and Gran plot.
Measurement modes: Auto-read, timed-read and continuous-read modes with Reading / Stable / Locked prompts.
Data capacity: up to 1000 data sets per parameter with GLP-oriented data functions.
Expansion note: optional DO module supports barometric and manual salinity compensation where selected.
Core technical summary
ParameterSpecification
ModelVIKING-ISE500
SKUPRO2561
Instrument typeBenchtop pH / ion / ORP / mV / temperature meter
Display7.0″ color touchscreen
pH range-2.000 to 20.000 pH
pH accuracy±0.002 pH
ISE range1.00E-9 to 9.99E9
ORP / mV range-2000.00 to 2000.00 mV
Data logging1000 data sets per parameter · GLP functions
ProtectionIP54
pH accuracy±0.002 pH
ISE calibration1–6 points
Data memory1000 sets / parameter
ConnectivityUSB / RS-232
Page 2 / 5
📊 Technical Specifications
pH, ISE, ORP / mV and temperature performance
ParameterSpecification
pH range-2.000 to 20.000 pH
pH resolution0.1 / 0.01 / 0.001 pH
pH accuracy±0.002 pH
pH calibration1–6 points; NIST / DIN / GB buffer groups; custom buffer group support
ISE range1.00E-9 to 9.99E9; 4 significant digits
ISE accuracy±0.3%
ISE calibration1–6 points
ISE methodsDirect reading, standard addition, sample addition and Gran plot
ISE unitsµg/L, mg/L, g/L, mmol/L and pX
pX range-2.000 to 20.000 pX
pX accuracy±0.002 pX
ORP / mV range-2000.00 to 2000.00 mV
ORP / mV resolution0.1 / 0.01 mV
ORP / mV accuracy±0.03% FS
Temperature range-5 to 130°C
Temperature resolution0.1°C
Temperature accuracy±0.1°C
Temperature compensationATC / MTC
System, data and connectivity
ParameterSpecification
Display7.0″ color touchscreen.
Operating systemIntelligent system with user, electrode, method, calibration and data management.
Reading modesAuto-read Fast / Medium / Slow, timed-read and continuous-read.
Reading promptsReading / Stable / Locked status prompts.
Electrode diagnosticsSlope and offset display, calibration reminder and slope-limit alert.
Data capacity1000 data sets per parameter.
GLP supportData review, comparison, recalculation and traceability-oriented data functions.
Optional module noteDO module available depending on configuration; barometric and manual salinity compensation supported when selected.
ConnectivityUSB and RS-232.
Protection ratingIP54.
Power supplyAC adapter, 100–240 V input / DC 24 V output.
Dimensions280 × 280 × 130 mm.
WeightApprox. 2500 g / 5.5 lb.
Expected HTMLOpen VIKING-ISE500 HTML specification
Expected PDFOpen VIKING-ISE500 PDF datasheet
Specification note
Electrode selection: final ISE electrode, reference electrode and standard solution set must match the target ion and sample matrix.
Calibration: confirm calibration points, ionic strength adjustment, buffer group and reporting unit according to the laboratory method.
Configuration: optional modules, electrodes and accessories should be confirmed in the manufacturer-approved quotation.
Technical values are provided for product selection and orientation. Final delivered configuration, electrode type, accessory set, optional module selection and documentation should be confirmed through the official COLO.Science quotation or manufacturer-confirmed offer.
Page 3 / 5
🧰 Standard Delivery, Configuration and Visual Support
VIKING-ISE500 benchtop pH ion meter with touchscreen
VIKING-ISE500 · benchtop touchscreen pH / ISE / ORP / mV / temperature meter
Typical standard delivery / configuration-dependent package
VIKING-ISE500 main unit: benchtop touchscreen analyzer, SKU PRO2561.
pH / ISE measurement module: installed or selected according to the ordered configuration.
Electrode holder and power adapter: bench setup accessories for routine measurement work.
Calibration solutions / starter set: provided according to selected package and application; ion standards depend on target ion.
USB cable: for PC / printer connection where selected or included.
User documentation: multilingual user manual and quick-start guidance.
Product image references
VIKING-ISE500 benchtop ion meter main unit
Main unit · 7.0″ touchscreen benchtop pH / ion meter
VIKING-ISE500 water quality laboratory analyzer SKU PRO2561
Specification path · WaterQualityP / VIKING-ISE500

Configuration notice — delivery depends on selected options

pH / ISE meters may be delivered in different configurations depending on the application, selected electrodes, optional modules, accessories and regional package. This document is for general technical orientation only. The exact delivered configuration must be confirmed in the official quotation and manufacturer-approved specification.

📌 Selection and Use Notes
For ISE measurement
Target ion: define the ion, expected range, required unit and sample matrix before selecting electrode and standards.
Method choice: direct reading, standard addition, sample addition or Gran plot should follow the validated laboratory procedure.
For pH / ORP workflows
pH calibration: select NIST, DIN, GB or custom buffers according to reporting and laboratory SOP requirements.
ORP / mV: confirm electrode type and maintenance procedure where millivolt readings are used for process or water monitoring.
For GLP workflows
Traceability: use user, method, calibration and data management functions for controlled measurement records.
Review: stored data sets can support comparison, review and recalculation depending on selected workflow.
Page 4 / 5
COLO.Science · Final Support Page
Manufacturer and support
COLO Lab Experts
Polje ob Sotli 4, SI-3255, Slovenia
Selection guidance: Send the target ion, expected concentration range, sample matrix, required reporting unit, electrode preference, pH / ORP requirements and any GLP or data-export requirements. COLO.Science can help confirm the correct VIKING-ISE500 configuration and accessory set.
COLO.Science · VIKING-ISE500 benchtop pH / ion meter support

Need help selecting the correct pH / ISE measurement configuration?

Send the target ion, expected concentration range, sample type, calibration approach, electrode type and reporting requirement. COLO.Science can help confirm whether VIKING-ISE500 is the correct platform for your laboratory, water-quality or method-development workflow.

Technical specification path: VIKING-ISE500 HTML technical specification
Official configuration and manufacturer-confirmed specification notice:
This technical specification is provided for orientation, product selection and general information only. It does not represent the final binding technical specification, delivered configuration, accessory set, procurement requirement or acceptance criterion for a specific unit. The final official technical specification is only the specification confirmed by the manufacturer and issued for the exact configuration through an official COLO.Science quotation, proforma invoice, contract document, order confirmation or manufacturer-approved technical offer. Any values, options, accessories or configurations shown in this document must be verified for the specific delivery and should not be used as an exclusion or elimination criterion in procurement procedures without written manufacturer confirmation.