| instrument identity |
| instrument | VIKING-DO510 |
| model | DO510T |
| product number | PRO4953 |
| parameters | DO / Temp. (DO Saturation) |
| display | Color high-contrast LCD touchscreen, 4.3″ |
| intelligent operation | User management, calibration management, electrode management, method management, data management, log management |
| dissolved oxygen concentration (DO) |
| sensor type | Polarographic |
| range | 0.00 to 99.99 ppm |
| resolution | 0.01 ppm |
| accuracy |
0.00–50.00 ppm: ±0.10 ppm
50.0–99.99 ppm: ±0.50 ppm
|
| calibration reminder | Yes |
| calibration points | Air-saturated water or zero point |
| barometric compensation | Yes |
| manual salinity factor correction | Yes |
| selectable pressure unit | kPa, mbar, Torr, Atm |
| DO saturation |
| range | (0.0 to 600.0)% |
| resolution | 0.10% |
| accuracy | ±2.0% |
| temperature |
| range | -10 to 135 °C (14 to 275 °F) |
| unit | °C, °F |
| resolution | 0.1 |
| accuracy | ±0.1 °C |
| measurement workflow & data |
| reading modes | Auto Read (Fast, Medium, Slow), Timed, Continuous |
| reading prompts | Reading, Stable, Locked |
| auto-hold | Yes (locks measurement endpoint) |
| temp. compensation | ATC, MTC |
| data management | Data Storage 1000 Groups (GLP) |
| GLP features | Yes |
| data analysis | Review, compare and recalculate results |
| reset | Factory default restore |
| inputs, outputs & protection |
| DO with temp. probe input | 4-pin aviation connector |
| USB | USB 2.0 (flash memory device, PC, scanner) |
| backlight | Yes |
| auto shutdown | 300, 600, 1200, 1800, 3600 sec., off |
| IP rating | IP65 |
| date and time | Yes |
| power & physical |
| power | Rechargeable lithium battery, AC adapter (100–240 V AC input, DC 5 V output) |
| dimensions | 90 × 255 × 40 mm |
| weight | 500 g (1.1 lb) |
| topic | recommendation | note |
| probe preparation |
Inspect membrane condition, electrolyte and stability before logging results. |
Polarographic probes require correct membrane/electrolyte condition for stable results. |
| calibration |
Calibrate with air-saturated water for routine use; apply zero oxygen calibration when required. |
Calibration reminder supports traceability and SOP compliance. |
| compensation |
Use ATC (or MTC when needed); apply barometric pressure and salinity correction for best accuracy. |
Pressure and salinity influence DO and % saturation calculations. |
| data management |
Store GLP groups (1000) and export results via USB for reporting and documentation. |
Built-in review/compare/recalculate supports data verification. |