TITRO20 · Manual Potentiometric Titrator
Technical data sheet · Informational document · Controlled electronic document
Product page:
TITRO20 on colo.si
TITRO20 · technical specification
Manual titration platform for KF-ready moisture analysis, endpoint titration and electrochemical titration workflows
SET preset endpoint
MAT manual titration
Trace moisture 0.001% and less
pH 0.00–14.00
mV -1400 to +1400
— Practical titration system for routine and method-development laboratories
Product Overview
Application profile
Karl Fischer moisture analysis: configured for trace moisture determination where water content control is critical.
Acid-base titration: suitable for acidity, alkalinity and neutralization-capacity determinations.
Redox titration: supports millivolt endpoint workflows using the included platinum electrode.
Precipitation titration: supports halide applications using the included silver electrode.
Why this configuration matters
SET mode: preset endpoint titration for repeatable routine work, including KF endpoint workflows.
MAT mode: manual titration for method development, teaching and specialized titration procedures.
Complete accessory base: magnetic stirrer, reference electrode, platinum electrode, silver electrode, pH electrode and pH standards.
Compact platform: a practical manual system for laboratories that do not require fully automated method libraries.
Key features
Manual potentiometric titrator: designed for endpoint-based titration using electrode potential response.
Karl Fischer ready: supports KF moisture workflows for trace water analysis down to 0.001% and less.
pH measurement: 0.00 to 14.00 pH range with 0.01 pH resolution.
mV measurement: -1400 mV to +1400 mV range with 1 mV resolution.
pH calibration: 1–2 point pH calibration with NIST pH standard solutions.
Electrode package: reference, platinum, silver and pH composite electrode supplied as standard set.
Core technical summary
| Parameter | Specification |
| Model | TITRO20 |
| SKU | PRO6190 |
| Instrument type | Manual potentiometric titrator |
| Titration modes | SET preset endpoint / MAT manual titration |
| KF capability | Trace moisture from 0.001% and less |
| pH range | 0.00 to 14.00 pH |
| mV range | -1400 mV to +1400 mV |
| pH calibration | 1–2 point |
| Included stirrer | Magnetic stirrer PRO6175 |
| Standard set | Stirrer + electrodes + NIST pH standards |
Titration modesSET / MAT
KF-ready use0.001% moisture
pH range0.00–14.00 pH
Included kitStirrer + electrodes
Page 2 / 5
Technical Specifications
Instrument identity and measurement ranges
| Parameter | Specification |
| Model | TITRO20 |
| Product number / SKU | PRO6190 |
| Instrument type | Manual potentiometric titrator |
| Measured parameters | pH, mV / ORP, titration endpoint response |
| Primary titration modes | SET (preset endpoint titration), MAT (manual titration) |
| pH range | 0.00 to 14.00 pH |
| pH resolution | 0.01 pH |
| mV range | -1400 mV to +1400 mV |
| mV resolution | 1 mV |
| pH calibration | 1–2 point calibration |
| pH standards | NIST pH 4.01, 7.00, 10.01; 50 ml/vial |
| KF moisture context | Trace moisture determination from 0.001% and less |
Titration modes
| Mode | Technical note |
| SET | Preset endpoint titration; suitable for routine endpoint workflows including KF moisture analysis. |
| MAT | Manual titration mode; suitable for method development, teaching and specialized applications. |
| Endpoint principle | Manual potentiometric titration based on electrode potential or pH response. |
| Workflow profile | Best suited for practical laboratories requiring controlled manual operation rather than full automation. |
Included accessories and application fit
| Accessory / area | Specification / role |
| Magnetic stirrer | PRO6175; homogeneous mixing during titration. |
| Reference electrode | PRO6178; stable potential reference. |
| Platinum electrode | PRO6179; redox titration workflows. |
| Silver electrode | PRO6135; precipitation titrations and halide analysis. |
| pH composite electrode | Standard pH measurement and acid-base titration workflows. |
| NIST pH standards | pH 4.01, 7.00, 10.01; 50 ml/vial. |
| KF application area | Moisture-sensitive pharmaceutical, chemical, oil, solvent, polymer and cosmetic samples. |
| Acid-base application | Acidity, alkalinity and neutralization capacity. |
| Redox application | Oxidizing/reducing agents, peroxide and vitamin C type workflows. |
| Precipitation application | Chloride, bromide, iodide and related halide determinations. |
| Product page | Open TITRO20 product page |
| Expected PDF | Open TITRO20 PDF |
Specification note
KF configuration: final Karl Fischer reagent set, vessel configuration and accessories should be confirmed for the target sample type.
Electrode selection: endpoint reliability depends on electrode type, reagent system, sample matrix and operator workflow.
Quotation scope: the approved COLO.Science quotation defines the final delivered accessories and documentation.
Technical values are provided for product selection and orientation. Final delivered configuration, electrode type, accessory set, reagent compatibility and documentation should be confirmed through the official COLO.Science quotation or manufacturer-confirmed offer.
Page 3 / 5
COLO.Science · Final Support Page
COLO.Science · TITRO20 titration support
Need help confirming the correct TITRO20 configuration?
Send the sample matrix, expected endpoint, moisture range, titration type, electrode requirement and documentation needs.
COLO.Science can help confirm whether TITRO20 is the correct manual titration platform for your laboratory workflow.
Official configuration and manufacturer-confirmed specification notice:
This technical specification is provided for orientation, product selection and general information only. It does not represent the final binding technical specification, delivered configuration, accessory set, reagent compatibility, procurement requirement or acceptance criterion for a specific unit. The final official technical specification is only the specification confirmed by the manufacturer and issued for the exact configuration through an official COLO.Science quotation, proforma invoice, contract document, order confirmation or manufacturer-approved technical offer. Any values, options, accessories or configurations shown in this document must be verified for the specific delivery and should not be used as an exclusion or elimination criterion in procurement procedures without written manufacturer confirmation.
Page 5 / 5