COLO.Science · Technical Specification
VIKING-PH300
Benchtop pH Meter · pH / mV / Temperature · 5.7″ LCD · IP54
Reliable benchtop pH meter for routine measurements with 5.7″ high‑resolution LCD, Auto‑Read / Timed / Continuous modes, Auto‑Hold endpoint detection, ATC/MTC temperature compensation and 1–5 point pH calibration with standard buffer recognition (NIST, DIN, GB, USA).
pH / mV / °C 5.7″ LCD Auto‑Read Auto‑Hold ATC / MTC 5‑point calibration IP54
Document type: Technical Specification Sheet
Format: Landscape (Word/Outlook-ready)
VIKING-PH300 benchtop pH meter
Instrument
VIKING-PH300 · COLO.Science
Address (HQ)
Polje ob Sotli 4
SI-3255, Slovenia
Contact
COLO.Science logo VIKING-PH300 · Technical Specification
Document ID: COL-TEC-VIKING-PH300-EN
Version: Landscape
Informational technical sheet · Non-binding · Controlled electronic document

VIKING-PH300 · benchtop pH meter technical specification sheet

pH -2.00 → 20.00 mV -2000.0 → 2000.0 Temp -5 → 110°C 0.01 pH resolution IP54 — Routine pH / mV / temperature workflows

📋 general & performance specifications

instrument identity
modelVIKING-PH300
product number (SKU)PRO6084
brand / lineCOLO.Science · COLO Lab Experts
instrument typeBenchtop pH / mV / Temperature meter
intended useRoutine pH and mV measurements with temperature compensation in research, QC, education and environmental labs.
pH measurement
pH range-2.00 to 20.00 pH
resolutionselectable 0.1 / 0.01 pH
accuracy±0.01 pH
calibration pointsup to 5 (1‑5 points)
buffer groupsNIST, DIN, GB, USA (customizable)
standard recognitionautomatic, with user-defined buffers possible
electrode diagnosticsslope & offset display
mV measurement
mV range-2000.0 to 2000.0 mV
resolution0.1 mV
accuracy±0.3 mV or ±0.1% FS
temperature measurement
temperature range-5 to 110 °C / 23 to 230 °F
unit°C, °F
resolution0.1 °
accuracy±0.2 °C
measurement workflow & data
reading modesAuto‑Read (Fast/Medium/Slow), Timed, Continuous
reading promptsReading, Stable, Locked
endpoint functionAuto‑Hold (senses and locks measurement endpoint)
temperature compensationATC, MTC
data storage500 data sets (GLP‑compliant)
resetfactory reset
inputs, display & protection
pH electrode inputBNC (Q9)
temperature probe input4‑pin aviation connector
display5.7″ high‑resolution LCD with backlight
auto shutdown1‑60 min programmable, or off
IP ratingIP54
outputsUSB (PC), RS‑232 (printer)
power & physical
powerAC Adapter, 100–240 V AC input, DC 9 V output
dimensions242 × 195 × 68 mm
weight900 g (1.98 lb)

⚙️ key instrument features

5.7″ high‑resolution LCDClear display with backlight for easy reading.
Multi‑point calibration (1‑5)NIST/DIN/GB/USA buffers + custom; slope & offset display.
Flexible reading modesAuto‑Read, Timed, Continuous with Auto‑Hold endpoint detection.
GLP data management500 sets with logging, review, export via USB/RS‑232.
Electrode diagnosticsSlope and offset display for condition monitoring.
USB & RS‑232Connect to PC or printer for data export.
IP54Splash protection for wet‑bench use.

🧪 typical laboratory applications

routine water quality analysis (pH with temperature compensation)
education and training labs (electrochemistry basics)
food & beverage QC (pH of raw materials and finished products)
general ORP / redox measurements using mV mode
buffer preparation and routine pH verification
environmental monitoring and field sample analysis (lab-based)

📦 included / typical delivery (reference)

  • VIKING-PH300 benchtop pH meter
  • E-301-QC 3‑in‑1 pH composite electrode
  • NIST pH standard solutions: 4.01 / 7.00 / 10.01 (50 mL each)
  • COLO‑5 electrode stand
  • AC adapter (100–240 V input, 9 V output)

🧩 options & notes

buffers & accessories Full range of NIST/DIN/USA/GB buffers, storage solutions, cleaning kits available.
electrode selection Choose from COLO’s wide portfolio for specific sample types (aqueous, viscous, high solids, etc.).
connectivity Optional data transfer software for PC connection (RS‑232 / USB).
calibration practice Use fresh buffers and maintain temperature consistency for best accuracy.

🔬 practical operation notes

topicrecommendationnote
calibration Use 2‑5 points for critical work; verify slope/offset trend over time NIST/DIN/GB/USA buffer recognition supported; keep buffers uncontaminated
temperature Use ATC when available; otherwise keep sample and buffers at similar temperature ATC/MTC supported
electrode care Rinse between samples; store electrode properly after use Maintenance directly impacts stability and drift
data export Use USB or RS‑232 for printing or PC transfer GLP‑compliant reports

📄 reference PDF: VIKING-PH300.pdf
🧪 application guidance: workflow‑based electrode & buffer selection
🛠️ maintenance: electrode cleaning & periodic inspection
🧾 procurement: binding spec = quotation package

📌 Contact & Support

Headquarters address
Polje ob Sotli 4
SI-3255, Slovenia
Application guidance: tell us your sample type and throughput (routine QC, education, field transfer).
We will propose the matching instrument configuration and electrodes for your workflow.
Quick contact
Web: colo.si
PDF Tech Spec: VIKING-PH300.pdf
Email: sales@colo.si
Knowledge hub: COLO.Science
Manufacturer & configuration notice (important):
Specifications in this document are provided for general reference and may represent a typical configuration. COLO reserves the right to change technical specifications without notice.