VIKING-M500 is a touchscreen benchtop multi-parameter analyzer for laboratory measurement of pH, conductivity (EC), ion concentration (ISE), dissolved oxygen (DO) and temperature-related electrochemical parameters.
The system combines a 7.0″ high-contrast color LCD touchscreen, intelligent operating software, GLP-compliant data storage and guided measurement workflows for water laboratories, environmental monitoring, industrial quality control and research applications.
| Range | −2.000 to 20.000 pH |
|---|---|
| Resolution | 0.1, 0.01, 0.001 pH |
| Accuracy | ±0.002 pH |
| Calibration points | Up to 6 |
| Standard customization | Yes |
| Calibration reminder | Yes |
| Standard recognition | DIN, NIST, USA, MERK, JIS, GB |
| Slope limit | Yes |
| Electrode diagnostics | Slope and offset display |
| Range | −2000.00 to 2000.00 mV |
|---|---|
| Resolution | 0.1, 0.01 mV |
| Accuracy | ±0.1 mV or ±0.03% of reading, whichever is greater |
| EH ORP mode | Yes |
| ORP calibration | 1 custom point, relative mV |
| Range | −2.000 to 20.000 pX |
|---|---|
| Resolution | 0.1, 0.01, 0.001 pX |
| Accuracy | ±0.002 pX |
| Calibration points | Up to 6 |
| Range | 1.000e−9 to 9.999e+9 |
|---|---|
| Units | mol/L, mmol/L, g/L, mg/L, μg/L, ppm, ppb |
| Resolution | Up to 4 significant digits |
| Accuracy | ±0.3% |
| Calibration points | Up to 6 |
| Modes | Direct Reading, Standard Addition, Sample Addition, GRAN |
| Built-in methods | F−, Cl−, Br−, I−, NO₃−, BF₄−, NH₄⁺, K⁺, Na⁺, Ca²⁺, Cu²⁺, Pb²⁺, Ag⁺ and user methods |
| Conductivity range | 0.000 μS/cm to 2000 mS/cm |
|---|---|
| Resolution | Minimum 0.001 μS/cm, range-dependent |
| Accuracy | ±0.5% FS |
| Reference temperature | 5, 10, 15, 18, 20, 25 °C |
| Calibration | Up to 5 points |
| Calibration reminder | Yes |
| Standard recognition | 10 μS/cm, 84 μS/cm, 500 μS/cm, 1413 μS/cm, 12.88 mS/cm; 146.5 μS/cm, 1408 μS/cm, 12.85 mS/cm, 111.3 mS/cm |
| Compensation | None, linear, pure water |
| User settings | Cell constant, temperature coefficient, TDS factor |
| Resistivity | 5.00 Ω·cm to 100.0 MΩ·cm |
|---|---|
| Resistivity resolution | Minimum 0.01 Ω·cm |
| TDS range | 0.000 to 1000 ppt |
| TDS resolution | Minimum 0.001 ppm, range-dependent |
| Salinity type | %NaCl |
| Salinity range | 0.0 to 80.0 ppt |
| Salinity resolution | 0.1 ppt |
| Salinity accuracy | ±1 ppt |
| Concentration range | 0.00 to 99.99 ppm |
|---|---|
| Concentration resolution | 0.01 ppm |
| Concentration accuracy | 0.00–50.00 ppm: ±0.10 ppm; 50.0–99.99 ppm: ±0.50 ppm |
| % saturation range | 0.0 to 600% |
| % saturation resolution | 0.10% |
| % saturation accuracy | ±2.0% |
| Calibration | Air-saturated water or zero point |
| Pressure compensation | Automatic barometric pressure compensation |
| Pressure units | kPa, mbar, Torr, Atm |
| Salinity correction | Manual factor correction |
| Calibration reminder | Yes |
| Temperature range | −10 to 135 °C / 14 to 275 °F |
|---|---|
| Units | °C, °F |
| Resolution | 0.1 |
| Accuracy | ±0.1 °C |
| Reading modes | Auto (Fast/Medium/Slow), Timed, Continuous |
| Prompts | Reading, Stable, Locked |
| Temp. compensation | ATC, MTC |
| Data storage | 1000 groups per parameter |
|---|---|
| GLP features | Yes |
| Log management | Yes |
| Data functions | Review, comparison and recalculation |
| pH electrode input | BNC (Q9) |
|---|---|
| DO + temperature probe | 4-pin aviation connector |
| Conductivity + temperature probe | 5-pin aviation connector |
| USB | USB 2.0 flash device; PC; scanner |
| RS-232 | Printer; auto-sampler |
| Display | 7.0″ color LCD touchscreen with backlight |
|---|---|
| Auto shutdown | 1–60 min or off |
| IP rating | IP54 |
| Date and time | Yes |
| Power | AC adapter, 100–240 V AC input, 24 V DC output |
220 × 195 × 68 mm
Main unit dimensions
950 g · 2.09 lb
Instrument body
USB / RS-232
PC, printer, scanner and auto-sampler workflows
Electrode and probe selection should be matched to the sample matrix, measurement range, required traceability and connector configuration. For complete system configuration, verify the supplied electrodes, optional ISE electrodes, calibration standards and consumables before purchase.
COLO.Science can support selection of the main unit, electrodes, probes, calibration standards, ISE methods and accessories for water quality, environmental, industrial QC and research laboratory applications.
VIKING-M500 is a benchtop multi-parameter analyzer for pH, EC, ISE, dissolved oxygen and temperature measurement with 7.0″ touchscreen, GLP data management, USB / RS-232 communication and IP54 housing.