COLO.Science · Technical Specification
VIKING-DO310
Portable DO Meter · DO/Temp. (DO Saturation) · Polarographic · IP65
Portable dissolved oxygen meter for field and laboratory measurements with polarographic DO sensing. Supports DO concentration (ppm), DO saturation (%) and temperature with a 3.5″ LCD display. Multi-reading workflow supports Auto-Read (Fast/Medium/Slow), Timed and Continuous, with Auto-Hold endpoint locking. Includes GLP-compliant data storage, USB communication, ATC/MTC, air-saturated water or zero oxygen calibration, automatic barometric pressure compensation, and manual salinity factor correction. Robust IP65 waterproof housing supports outdoor and field measurements.
DO 0.00–50.00 ppm DO SAT 0.0–300% TEMP -5 → 110°C POLAROGRAPHIC 3.5″ LCD ATC / MTC GLP USB IP65
Document type: Technical Specification Sheet
Format: Landscape (Word/Outlook-ready)
VIKING-DO310 portable dissolved oxygen meter
Instrument
VIKING-DO310 · COLO.Science
Address (HQ)
Polje ob Sotli 4
SI-3255, Slovenia
Contact
COLO.Science logo VIKING-DO310 · Technical Specification
Document ID: COL-TEC-VIKING-DO310-EN
Version: Landscape
Informational technical sheet · Non-binding · Controlled electronic document

VIKING-DO310 · Portable DO Meter technical specification sheet

DO 0.00–50.00 ppm DO SAT 0.0–300% TEMP -5 → 110°C POLAROGRAPHIC 3.5″ LCD USB IP65 — Auto-Hold / Auto-Read + Timed + Continuous / pressure & salinity correction

General & performance specifications

instrument identity
instrumentVIKING-DO310
modelDO310F
product numberPRO3164
parametersDO / Temp. (DO Saturation)
displayLCD display screen, 3.5″
general features (workflow)
multi-readingAuto-Read (Fast, Medium, Slow), Timed, Continuous
auto-holdYes (senses and locks the measurement endpoint)
temp. compensationAutomatic / Manual temperature compensation (ATC, MTC)
USB communicationYes (PC, printer)
auto power-offYes (extends battery service life)
resetFactory default restore
IP ratingIP65 waterproof
dissolved oxygen concentration (DO)
sensor typePolarographic
range0.00 to 50.00 ppm
resolution0.01 ppm
accuracy±0.1 ppm
calibration pointsAir-saturated water or zero point
barometric compensationYes (automatic)
manual salinity factor correctionYes
selectable pressure unitkPa, mbar, Torr, Atm
DO saturation
range(0.0 to 300.0)%
resolution0.10%
accuracy±2.0%
temperature
range-5 to 110 °C (23 to 230 °F)
unit°C, °F
resolution0.1
accuracy±0.2 °C
measurement workflow & data
reading promptsReading, Stable, Locked
data managementData storage (GLP) — up to 1000 groups
GLP featuresYes
inputs, outputs & protection
DO with temp. probe input4-pin aviation connector
outputsUSB (PC, printer)
backlightYes
auto shutdown300, 600, 1200, 1800, 3600 sec., off
date and timeYes
power & physical
powerRechargeable lithium battery, AC adapter (100–240 V AC input, DC 5 V output)
dimensions80 × 225 × 35 mm
weight400 g (0.88 lb)
dimensions (carton)410 × 330 × 165 mm
weight (gross)5.5 kg

Key instrument features

3.5″ LCD displayClear screen for field operation with stable/locked prompts.
Multi-reading modesAuto-Read (Fast/Medium/Slow), Timed and Continuous modes for SOP-aligned workflows.
Auto-Hold endpointSenses and locks the measurement endpoint to reduce user variability.
Compensation toolsATC/MTC, automatic barometric pressure compensation, manual salinity factor correction.
GLP data integrityGLP-compliant storage for traceable documentation and reporting.
USB communicationUSB interface for communication with PC and printer.
IP65Waterproof housing suitable for outdoor and field measurements.

Typical applications

field dissolved oxygen measurements (rivers, lakes, reservoirs)
aquaculture and fisheries monitoring (DO and % saturation control)
wastewater and process water monitoring
environmental and municipal water laboratories
spot checks and verification in QC workflows

Included / typical delivery (reference)

  • PRO6163 polarographic DO electrode kit (zero oxygen calibration solution, dissolved oxygen electrolyte, 3 pcs electrode membrane cover)
  • Electrode holder
  • Silicone rubber case
  • Wristbands
  • Carrying case

Options & notes

calibration practice Use air-saturated water for routine calibration; use a zero point when SOP/validation requires.
pressure & salinity Set the correct pressure unit and apply salinity correction for saline samples to maintain accuracy.
power management Use Auto Shutdown to extend battery service life during fieldwork.
configuration notice Final delivered configuration may vary by region and quotation package; specifications may change without notice.

Practical operation notes

topicrecommendationnote
probe preparation Inspect membrane condition, electrolyte and stability before logging results. Polarographic probes require correct membrane/electrolyte condition for stable results.
calibration Calibrate with air-saturated water for routine use; apply zero oxygen calibration when required. Supports air-saturated water or zero point calibration.
compensation Use ATC (or MTC when needed); apply barometric pressure and salinity correction for best accuracy. Pressure and salinity influence DO and % saturation calculations.
data management Store GLP groups and export results via USB for reporting and documentation. USB supports PC/printer connection for documentation workflows.

Contact & Support

Headquarters address
Polje ob Sotli 4
SI-3255, Slovenia
Application guidance: tell us your environment (freshwater, wastewater, process water, aquaculture), the required reporting format (ppm, % saturation, or both) and typical temperature range. We will propose the optimal configuration and accessories for your workflow.
Quick contact
Web: colo.si
PDF Tech Spec: VIKING-DO310.pdf
Email: sales@colo.si
Knowledge hub: COLO.Science
Manufacturer & configuration notice (important):
Specifications in this document are provided for general reference and may represent a typical configuration. COLO.Science reserves the right to change technical specifications without notice.

For procurement and tender purposes, the only valid and binding specification is the official COLO.Science quotation and the accompanying technical specification that is engineering-approved.