COLO.Science · Technical Specification
VIKING-DO310 · Technical Specification
| instrument identity | |
|---|---|
| instrument | VIKING-DO310 |
| model | DO310F |
| product number | PRO3164 |
| parameters | DO / Temp. (DO Saturation) |
| display | LCD display screen, 3.5″ |
| general features (workflow) | |
| multi-reading | Auto-Read (Fast, Medium, Slow), Timed, Continuous |
| auto-hold | Yes (senses and locks the measurement endpoint) |
| temp. compensation | Automatic / Manual temperature compensation (ATC, MTC) |
| USB communication | Yes (PC, printer) |
| auto power-off | Yes (extends battery service life) |
| reset | Factory default restore |
| IP rating | IP65 waterproof |
| dissolved oxygen concentration (DO) | |
| sensor type | Polarographic |
| range | 0.00 to 50.00 ppm |
| resolution | 0.01 ppm |
| accuracy | ±0.1 ppm |
| calibration points | Air-saturated water or zero point |
| barometric compensation | Yes (automatic) |
| manual salinity factor correction | Yes |
| selectable pressure unit | kPa, mbar, Torr, Atm |
| DO saturation | |
| range | (0.0 to 300.0)% |
| resolution | 0.10% |
| accuracy | ±2.0% |
| temperature | |
| range | -5 to 110 °C (23 to 230 °F) |
| unit | °C, °F |
| resolution | 0.1 |
| accuracy | ±0.2 °C |
| measurement workflow & data | |
| reading prompts | Reading, Stable, Locked |
| data management | Data storage (GLP) — up to 1000 groups |
| GLP features | Yes |
| inputs, outputs & protection | |
| DO with temp. probe input | 4-pin aviation connector |
| outputs | USB (PC, printer) |
| backlight | Yes |
| auto shutdown | 300, 600, 1200, 1800, 3600 sec., off |
| date and time | Yes |
| power & physical | |
| power | Rechargeable lithium battery, AC adapter (100–240 V AC input, DC 5 V output) |
| dimensions | 80 × 225 × 35 mm |
| weight | 400 g (0.88 lb) |
| dimensions (carton) | 410 × 330 × 165 mm |
| weight (gross) | 5.5 kg |
| 3.5″ LCD display | Clear screen for field operation with stable/locked prompts. |
| Multi-reading modes | Auto-Read (Fast/Medium/Slow), Timed and Continuous modes for SOP-aligned workflows. |
| Auto-Hold endpoint | Senses and locks the measurement endpoint to reduce user variability. |
| Compensation tools | ATC/MTC, automatic barometric pressure compensation, manual salinity factor correction. |
| GLP data integrity | GLP-compliant storage for traceable documentation and reporting. |
| USB communication | USB interface for communication with PC and printer. |
| IP65 | Waterproof housing suitable for outdoor and field measurements. |
| field dissolved oxygen measurements (rivers, lakes, reservoirs) |
| aquaculture and fisheries monitoring (DO and % saturation control) |
| wastewater and process water monitoring |
| environmental and municipal water laboratories |
| spot checks and verification in QC workflows |
| calibration practice | Use air-saturated water for routine calibration; use a zero point when SOP/validation requires. |
| pressure & salinity | Set the correct pressure unit and apply salinity correction for saline samples to maintain accuracy. |
| power management | Use Auto Shutdown to extend battery service life during fieldwork. |
| configuration notice | Final delivered configuration may vary by region and quotation package; specifications may change without notice. |
| topic | recommendation | note |
|---|---|---|
| probe preparation | Inspect membrane condition, electrolyte and stability before logging results. | Polarographic probes require correct membrane/electrolyte condition for stable results. |
| calibration | Calibrate with air-saturated water for routine use; apply zero oxygen calibration when required. | Supports air-saturated water or zero point calibration. |
| compensation | Use ATC (or MTC when needed); apply barometric pressure and salinity correction for best accuracy. | Pressure and salinity influence DO and % saturation calculations. |
| data management | Store GLP groups and export results via USB for reporting and documentation. | USB supports PC/printer connection for documentation workflows. |