VIKING-DO210
Portable DO Meter · DO/Temp. (DO Saturation) · Polarographic · IP65
Portable dissolved oxygen meter for field and laboratory measurements with polarographic DO sensing. Supports DO concentration (mg/L), DO saturation (%) and temperature with a 3.5″ LCD display. Measurement workflow supports Auto Read and Continuous modes, with Auto-Hold endpoint locking. Includes GLP-style data storage (200 results per parameter), ATC/MTC, support for air-saturated water or zero oxygen calibration, manual barometric pressure compensation and manual salinity factor correction. Robust IP65 waterproof housing supports outdoor and field measurements.
DO 0.00–20.00 mg/L DO SAT 0.0–200% TEMP -5 → 110°C POLAROGRAPHIC 3.5″ LCD ATC / MTC DATA 200 IP65
Document type: Technical Specification Sheet
Format: Landscape (Word/Outlook-ready)
VIKING-DO210 portable dissolved oxygen meter
Instrument
VIKING-DO210 · COLO.Science
Address (HQ)
Polje ob Sotli 4
SI-3255, Slovenia
Contact
COLO.Science logo VIKING-DO210 · Technical Specification
Document ID: COL-TEC-VIKING-DO210-EN
Version: Landscape
Informational technical sheet · Non-binding · Controlled electronic document

VIKING-DO210 · Portable DO Meter technical specification sheet

DO 0.00–20.00 mg/L DO SAT 0.0–200% TEMP -5 → 110°C POLAROGRAPHIC 3.5″ LCD IP65 — Auto-Hold / Auto Read + Continuous / manual pressure & salinity correction

General & performance specifications

instrument identity
instrumentVIKING-DO210
modelDO210E
product numberPRO3316
parametersDO / Temp. (DO Saturation)
displayLCD display screen, 3.5″
general features (workflow)
multi-readingAuto Read, Continuous
auto-holdYes (senses and locks the measurement endpoint)
temp. compensationAutomatic / Manual temperature compensation (ATC, MTC)
data capacity200 results for each parameter
auto power-offYes (extends battery service life)
resetFactory default restore
IP ratingIP65 waterproof
dissolved oxygen concentration (DO)
sensor typePolarographic
range0.00 to 20.00 mg/L
resolution0.01 mg/L
accuracy±0.3 mg/L
calibration pointsAir-saturated water or zero point
barometric compensationYes (manual)
manual salinity factor correctionYes
DO saturation
range(0.0 to 200.0)%
resolution0.1%
accuracy±2.0%
temperature
range-5 to 110 °C
unit°C
resolution0.1 °C
accuracy±0.2 °C
measurement workflow & data
reading promptsReading, Stable, Locked
data managementData storage: 200 results each
temp. compensationATC, MTC
inputs, outputs & protection
DO with temp. probe input4-pin aviation connector
backlightYes
auto shutdown300, 600, 1200, 1800, 3600 sec., off
IP ratingIP65
power & physical
powerRechargeable lithium battery, AC adapter (100–240 V AC input, DC 5 V output)
dimensions80 × 255 × 35 mm
weight400 g (0.88 lb)

Key instrument features

3.5″ LCD displayClear screen for field operation with Reading/Stable/Locked prompts.
Reading modesAuto Read and Continuous for routine DO monitoring.
Auto-Hold endpointSenses and locks the measurement endpoint to reduce user variability.
Compensation toolsATC/MTC, manual barometric pressure compensation, manual salinity factor correction.
Data capacityUp to 200 results stored for each parameter (DO / % saturation / temperature).
IP65Waterproof housing suitable for outdoor and field measurements.

Typical applications

field dissolved oxygen checks (rivers, lakes, reservoirs)
aquaculture and fisheries monitoring (DO and % saturation control)
wastewater and process water monitoring
environmental and municipal water laboratories
spot checks and verification in QC workflows

Included / typical delivery (reference)

  • PRO6162 polarographic DO electrode kit (zero oxygen calibration solution, dissolved oxygen electrolyte, 3 pcs electrode membrane cover)
  • Electrode holder
  • Silicone rubber case
  • Wristbands

Options & notes

calibration practice Use air-saturated water for routine calibration; use a zero point when SOP/validation requires.
pressure & salinity Apply manual pressure compensation and salinity correction for more reliable reporting across conditions.
power management Use Auto Power-Off to extend battery service life during fieldwork.
configuration notice Final delivered configuration may vary by region and quotation package; specifications may change without notice.

Practical operation notes

topicrecommendationnote
probe preparation Inspect membrane condition and electrolyte level before logging results. Polarographic probes require correct membrane/electrolyte condition for stable results.
calibration Calibrate with air-saturated water for routine use; apply zero oxygen calibration when required. Supports air-saturated water or zero point calibration.
compensation Use ATC (or MTC when needed); apply manual pressure and salinity correction for best accuracy. Pressure and salinity influence DO and % saturation calculations.
data management Store results (200 per parameter) and log stable endpoints using Auto-Hold. Reading prompts (Reading/Stable/Locked) support consistent field sampling.

Contact & Support

Headquarters address
Polje ob Sotli 4
SI-3255, Slovenia
Application guidance: tell us your environment (freshwater, wastewater, process water, aquaculture), the required reporting format (mg/L, % saturation, or both) and typical temperature range. We will propose the optimal configuration and accessories for your workflow.
Quick contact
Web: colo.si
PDF Tech Spec: VIKING-DO210.pdf
Email: sales@colo.si
Knowledge hub: COLO.Science
Manufacturer & configuration notice (important):
Specifications in this document are provided for general reference and may represent a typical configuration. COLO.Science reserves the right to change technical specifications without notice.

For procurement and tender purposes, the only valid and binding specification is the official COLO.Science quotation and the accompanying technical specification that is engineering-approved.
```