COLO.Science · Technical Specification
FUMECAB70 · Ducted Fume Hood · Technical Specification
| Parameter | Specification |
|---|---|
| Model | FUMECAB70 |
| Internal size (W×D×H) | 640 × 550 × 700 mm |
| External size (W×D×H) | 700 × 620 × 1150 mm |
| Maximum opening | 660 mm |
| Air flow velocity | 0.3 – 0.8 m/s (6 speeds) |
| Pre‑filter | Polyester fiber, washable |
| Main filter | Active carbon filter |
| Noise level | ≤ 60 dB(A) |
| Front window | Manual, acrylic material |
| UV lamp | Optional |
| LED lamp | 2 pcs × 4W |
| Power consumption | 400 W |
| Display | LED (fan speed indication) |
| Work surface material | Chemical resistant phenolic resin |
| Main body material | Cold rolled steel with anti‑bacterial coating |
| Power supply | 220 V, 50/60 Hz |
| Standard configuration | LED Lamp ×2, Active Carbon filter, 4 meters exhaust PVC duct, Pipe strap |
| Optional configuration | Base stand, UV Lamp |
| Main body | Cold‑rolled steel with anti‑bacterial coating, resistant to corrosion and easy to clean |
| Work surface | Chemical‑resistant phenolic resin – withstands acids, solvents, and spills |
| Front window | Acrylic, manual sliding, provides excellent visibility |
| Filtration system | Washable polyester pre‑filter + active carbon main filter |
| Fan | 6‑speed adjustable, airflow 0.3–0.8 m/s |
| Lighting | 2 × 4W LED lamps inside the work area |
| Recommended application | Safe handling of chemical vapors, fumes, and odors in laboratories, teaching labs, and research facilities |
| Filter monitoring | Replace active carbon filter periodically based on usage and chemical exposure |
| Cleaning | Wipe work surface and exterior with mild detergent; avoid abrasive cleaners |
| Environment | Indoor use, well‑ventilated area, temperature 5–40°C |