COLO.Science · Technical Specification
FLUO-IN900 · Scientific Research Inverted Fluorescence Microscope · Technical Specification
| Parameter | Specification / Value |
|---|---|
| Microscope type | FLUO-IN900 scientific research inverted fluorescence microscope |
| Observation modes | Bright field, phase contrast, fluorescence, polarized light and DIC-ready configuration; the series description also positions the platform for advanced life science imaging workflows |
| Optical system | Unique NIS infinite optical system |
| Eyepieces | HWF 10× / 25 mm eyepieces |
| Viewing head | Trinocular head with 45° inclined tubes |
| Beam splitter positions | 100-0 / 20-80 / 0-100 |
| Interpupillary distance | 47-78 mm |
| Diopter adjustment | Adjustable on both eyepieces |
| Nosepiece | Reversed sextuple nosepiece on ball bearings |
| DIC compatibility | Optional DIC prisms for 10×, 20× and 40× objectives |
| Optical treatment | Objectives are anti-fungus treated and anti-reflection coated for high light throughput |
| Standard objective 1 | Infinity Plan Semi-Apochromatic Phase PH 10× / NA 0.30 / WD 10.2 mm / cover glass correction 1.2 mm |
| Standard objective 2 | Infinity Plan Semi-Apochromatic Phase PH 20× / NA 0.45 / WD 7.5-8.8 mm with collar / cover glass correction 1.2 mm |
| Standard objective 3 | Infinity Plan Semi-Apochromatic Phase PH 40× / NA 0.60 / WD 3-4.4 mm with collar / cover glass correction 1.2 mm |
| Stage | Mechanical stage 340 × 230 mm with coaxial X-Y movement 130 × 85 mm |
| Included stage inserts | Metal insert for Ø38 mm culture dish; metal insert for Ø54 mm dish or 76 × 26 mm slides; metal insert for multiwell plates (127.76 × 85.48 mm); metal insert for Terasaki plates (83.3 × 58.0 mm) |
| Focusing | Coaxial coarse and fine adjustment with 200 graduations |
| Fine focus precision | 1 µm; 0.2 mm per rotation |
| Total focus travel | Approximately 10 mm |
| Condenser | Rotating condenser, N.A. 0.55 with iris diaphragm |
| Condenser working distance | 26 mm |
| Phase annuli | For 10×/20×, 40× and 60× objectives |
| Condenser DIC compatibility | DIC prisms for 10×/20×/40× and 60× objectives (additional accessories required) |
| Filter holders | Three positions for Ø38 mm filters |
| Condenser arm | Tilting arm for easy stage access and placement of larger items such as flasks and Petri dishes |
| Lens changer | Integrated 1.5× lens changer for intermediate magnifications without refocusing |
| Parameter | Specification / Value |
|---|---|
| Transmitted illumination | 10 W LED transmitted illumination with adjustable intensity |
| Power supply | Internal power supply, 100-240 V |
| Illumination style | Köhler LED transmitted illumination |
| Fluorescence light source | 100 W mercury vapor HBO lamp |
| Fluorescence attachment | Six-position fluorescence attachment with UV shield |
| Standard fluorescence filters | Blue and green interference Ø45 mm filters |
| Standard fluorescence filter set 1 | Blue excitation: EX 460-495 nm (BP), DM 505, EM 510 nm (LP) |
| Standard fluorescence filter set 2 | Green excitation: EX 510-550 nm (BP), DM 570, EM 575 nm (LP) |
| Optional filter set - blue | EX 460-490 nm (BP), DM 500 nm, EM 520 nm (LP), requires empty filter cube |
| Optional filter set - green | EX 510-550 nm (BP), DM 570 nm, EM 590 nm (LP), requires empty filter cube |
| Optional filter set - violet | EX 400-410 nm (BP), DM 455 nm, EM 460 nm (LP), requires empty filter cube |
| Optional filter set - ultra-violet | EX 330-385 nm (BP), DM 400 nm, EM 420 nm (LP), requires empty filter cube |
| Optional filter set - red | BP 620-650, dichroic mirror DM 660, barrier BA 670-750, requires empty filter cube |
| Additional fluorescence accessories | 100 W power supplier NFP-1, FL-ND6 filter and FL-ND25 filter are listed in the accessory range |
| Parameter | Specification / Value |
|---|---|
| Optional phase objective | 4× / NA 0.13 / WD 17.78 mm; requires phase annuli for 4× objective |
| Optional phase objective | 60× / NA 0.70 / WD 1.8-2.6 mm with collar / cover glass correction 0.1-1.3 mm |
| Optional fluorescent objective | 4× / NA 0.13 / WD 17 mm |
| Optional fluorescent objective | 10× / NA 0.30 / WD 7.3 mm / cover glass correction 1.2 mm |
| Optional fluorescent objective | 20× / NA 0.45 / WD 8 mm / cover glass correction 1.2 mm |
| Optional fluorescent objective | 40× / NA 0.60 / WD 3.6 mm / cover glass correction 1.3 mm |
| Optional fluorescent objective | 60× / NA 0.70 / WD 1.8-2.6 mm with collar / cover glass correction 0.1-1.3 mm |
| Optional Plan-Apo objective | 10× / NA 0.45 / WD 4.0 mm / cover glass correction 0.17 mm |
| Digital imaging option | 9.7" LCD screen integrated with 5 MP digital camera, Android OS, HDMI, USB, SD card, Wi-Fi and mouse control |
| C-mount accessory | 1× magnification C-mount for C-mount camera |
| C-mount relay option | 0.50× relay objective for 1/2" C-mount camera |
| C-mount relay option | 0.35× relay objective for 1/3" C-mount camera |
| C-mount relay option | 0.63× relay objective for 2/3" C-mount camera |
| SLR photography adapter | Universal Ø23.2 mm tube adapter with built-in 2× lens for APS-C SLR camera; T2 ring required |
| Camera T2 rings | T2 ring for Nikon D SLR digital camera; T2 ring for Canon EOS SLR digital camera |
| Additional holders | Universal cell culture holder and rotating magnification changer with 1.5× lens are listed as accessories |
| Base microscope set | FLUO-IN900 inverted microscope body, trinocular viewing head, HWF 10×/25 mm eyepieces, reversed sextuple nosepiece, long working distance phase objective set as listed in the overview, large mechanical stage and rotating N.A. 0.55 condenser |
| Illumination bundle | 10 W LED transmitted illumination with internal 100-240 V power supply, plus fluorescence module with 100 W HBO lamp |
| Fluorescence bundle | Six-position fluorescence attachment with UV shield and standard blue / green excitation filter configuration |
| Stage accessory bundle | Included inserts for Ø38 mm culture dish, Ø54 mm dish or 76 × 26 mm slides, multiwell plates and Terasaki plates |
| Configuration note | Several objectives, DIC components, camera modules, C-mounts and additional fluorescence filter sets are listed as optional accessories. Final supplied scope should follow quotation and order confirmation. |