COLO.SCIENCE | TECHNICAL SPECIFICATION

MBINT Series

Microbiological Incubators | 55, 90 and 165 L | RT+5 to 85 °C

MBINT microbiological incubators provide precise, uniform temperature conditions for cultivation, microbiological testing, cell-culture support and routine quality-control work. Forced-air circulation, ARM processor control and a TFT capacitive display support repeatable daily operation.

MBINT55MBINT90MBINT165±0.2 °C at 37 °CTFT touch displaySUS304 chamberInner glass door

Document type: Technical Specification Sheet
Product family: MBINT Microbiological Incubator Series
Models: MBINT55 / MBINT90 / MBINT165
Format: A4 landscape | COLO.Science TechSpec v1.1

Technical review note: the source cover states 0-100 °C, while the model table states RT+5 to 85 °C. This specification uses the model-table value pending final manufacturer confirmation.
COLO MBINT microbiological incubator
VOLUME55 / 90 / 165 L
RANGERT+5 to 85 °C
CONTACTsales@colo.si

MBINT | Microbiological Incubator Series

Technical data sheet | Informational document
Models: MBINT55 / MBINT90 / MBINT165

Controlled incubation with uniform forced-air heating

Series overview

Temperature control

  • ARM digital PID algorithm for temperature, time and process control.
  • Forced-air circulation for chamber temperature uniformity.
  • 1-, 2- or 3-point calibration listed in the source specification.
  • Heating to 37 °C in less than 15 minutes under stated source conditions.

Interface and workflow

  • TFT capacitive graphic display with backlight.
  • Timing range from 0 to 99.9 hours.
  • Door-open, overtemperature and system-instability alarms.
  • User-set temperature-limit monitoring.

Chamber and sample access

  • Corrosion-resistant SUS304 stainless-steel inner chamber.
  • Tempered inner glass door for observation with reduced heat loss.
  • Test port for external probes or validation work.
  • Two adjustable shelves supplied as standard.

Safety

  • Digital overtemperature control.
  • Independent analogue hydraulic thermostat.
  • Recommended overtemperature offset: 10%, according to the source.
  • Protection concept designed to reduce overheating risk to samples.
ACCURACY AT 37 °C
±0.2 °C
UNIFORMITY AT 37 °C
±0.2 °C
STANDARD SHELVES
2 adjustable

MBINT | Models and Technical Specifications

Source-table values
Verify final configuration before order

Model selection
ModelChamber volumeTemperature rangePower consumptionInternal chamber (W x D x H)External dimensions (W x D x H)Net weight
MBINT5555 LRT+5 to 85 °C345 W395 x 346 x 395 mm536 x 588 x 681 mm44 kg
MBINT9090 LRT+5 to 85 °C385 W500 x 364 x 500 mm650 x 573 x 793 mm51 kg
MBINT165165 LRT+5 to 85 °C495 W525 x 525 x 600 mm730 x 750 x 900 mm69 kg

Temperature performance and control

Timing range0-99.9 h
Accuracy±0.2 °C at 37.0 °C
Temperature uniformity±0.2 °C at 37.0 °C
Heating time to 37 °CLess than 15 min
ControllerARM processor with digital PID algorithm
Temperature calibration1-, 2- or 3-point
DisplayTFT capacitive graphic display with backlight

Construction, safety and utilities

Inner chamberSUS304 stainless steel
Chamber heatingForced-air circulation controlled by ARM processor
Inner doorTempered glass
Test portYes
Standard shelves2 pcs, adjustable
Overheating protectionDigital control plus independent analogue hydraulic thermostat
AlarmsOverheating, door open, unstable system, set-limit exceeded
Power supplyAC 220 V ±10%, 50/60 Hz
Model focus for current catalogue work: MBINT90, 90 L, Vtiger product PRO4805, supplier model BJPX-H88BK/D. Confirm nomenclature, power supply and final temperature range in the formal quotation.
COLO.SCIENCE | PRODUCT SUPPORT
MANUFACTURER AND SUPPORT

COLO Lab Experts

Polje ob Sotli 4, SI-3255, Slovenia

Selection guidance: send the target setpoint, ambient conditions, required chamber volume, sample load, validation needs, access-port requirements and preferred alarm workflow.

QUICK CONTACT

Web: colo.si

Email: sales@colo.si

Phone: +386 70 713 437

Series: MBINT55 / MBINT90 / MBINT165

COLO.SCIENCE | MBINT INCUBATOR SUPPORT

Need help selecting the right chamber size and configuration?

Send your application, required working temperature, sample quantity and dimensions, door-opening frequency, recovery-time expectation, validation procedure and any need for external probes. COLO.Science can confirm the model and approved configuration.

MAIN TEMPERATURE CONTROL NAVIGATION