MBINT microbiological incubators provide precise, uniform temperature conditions for cultivation, microbiological testing, cell-culture support and routine quality-control work. Forced-air circulation, ARM processor control and a TFT capacitive display support repeatable daily operation.
Document type: Technical Specification Sheet
Product family: MBINT Microbiological Incubator Series
Models: MBINT55 / MBINT90 / MBINT165
Format: A4 landscape | COLO.Science TechSpec v1.1

| Model | Chamber volume | Temperature range | Power consumption | Internal chamber (W x D x H) | External dimensions (W x D x H) | Net weight |
|---|---|---|---|---|---|---|
| MBINT55 | 55 L | RT+5 to 85 °C | 345 W | 395 x 346 x 395 mm | 536 x 588 x 681 mm | 44 kg |
| MBINT90 | 90 L | RT+5 to 85 °C | 385 W | 500 x 364 x 500 mm | 650 x 573 x 793 mm | 51 kg |
| MBINT165 | 165 L | RT+5 to 85 °C | 495 W | 525 x 525 x 600 mm | 730 x 750 x 900 mm | 69 kg |
| Timing range | 0-99.9 h |
| Accuracy | ±0.2 °C at 37.0 °C |
| Temperature uniformity | ±0.2 °C at 37.0 °C |
| Heating time to 37 °C | Less than 15 min |
| Controller | ARM processor with digital PID algorithm |
| Temperature calibration | 1-, 2- or 3-point |
| Display | TFT capacitive graphic display with backlight |
| Inner chamber | SUS304 stainless steel |
| Chamber heating | Forced-air circulation controlled by ARM processor |
| Inner door | Tempered glass |
| Test port | Yes |
| Standard shelves | 2 pcs, adjustable |
| Overheating protection | Digital control plus independent analogue hydraulic thermostat |
| Alarms | Overheating, door open, unstable system, set-limit exceeded |
| Power supply | AC 220 V ±10%, 50/60 Hz |