COLO.Science · Technical Specification
MBIND · Laboratory Incubator Family · Technical Specification
| Product designation | MBIND Standard Laboratory Incubator Family |
| Equipment type | Standard laboratory incubator family with natural air circulation and programmable temperature control |
| Typical fields of use | Microbial testing, bacteria cultivation, incubation of cell cultures for in vitro fertilization, stability testing in pharmacy, industrial laboratories and medicine research |
| Air circulation principle | Natural air circulation for high temperature uniformity in chamber and reduced risk of cross contamination |
| Temperature range | RT+5°C to 80°C |
| Family variants | MBIND45, MBIND55, MBIND65, MBIND125 and MBIND210 |
| Configuration note | The MBIND family is a natural-air laboratory incubator platform. Exact chamber size, power, energy consumption and optional accessories depend on selected model and final quotation. |
| Model | Working volume | Inner chamber size (W × D × H) | Outer dimensions (W × D × H) | Power | Energy consumption at 37°C |
|---|---|---|---|---|---|
| MBIND45 | 45 L | 350 × 350 × 350 mm | 525 × 480 × 620 mm | 0.35 kW | 20 Wh/h |
| MBIND55 | 53 L | 370 × 350 × 420 mm | 520 × 540 × 710 mm | 0.20 kW | 23 Wh/h |
| MBIND65 | 62 L | 400 × 350 × 440 mm | 614 × 577 × 625 mm | 0.45 kW | 25 Wh/h |
| MBIND125 | 120 L | 500 × 450 × 550 mm | 675 × 580 × 820 mm | 0.60 kW | 40 Wh/h |
| MBIND210 | 210 L | 600 × 580 × 600 mm | 775 × 710 × 870 mm | 0.70 kW | 60 Wh/h |
| Air circulation | Natural air circulation |
| Temperature range | RT+5°C to 80°C |
| Temperature setting accuracy | 0.1°C |
| Temperature variation at 37°C | ± 0.6°C |
| Temperature variation at 50°C | ± 1.0°C |
| Temperature fluctuation at 37°C | ± 0.2°C |
| Temperature fluctuation at 50°C | ± 0.3°C |
| Heat-up time | 25 min to 37°C and 40 min to 50°C for smaller variants; 30 min to 37°C and 40 min to 50°C for larger variants |
| Display | LED display with backlight showing current and set temperature and time, temperature unit (Celsius or Fahrenheit), programme time, time zones and summertime / wintertime |
| Temperature control mode | Two-temperature-section intelligent PID control with sensor deviation correction and auto tuning |
| Timer | Adjustable timing with wait function; MBIND45 listed with 0–9999 days timing range, other family variants listed with 1 minute to 99 days |
| Temperature sensor | Pt100 DIN class A in 4-wire circuit |
| Door construction | Double-layer door structure with inner tempered glass observation window and silicone packing for sample observation with minimal heat loss |
| Safety protection | Class 3.1 integrated independent temperature safety device according to DIN 12880, with overheating visual and sound alarm |
| Operation functions | Temperature setting, timing function, auto stop, internal parameter locking, temperature correction, auto tuning, parameter memory in case of electrical cut-off, over-temperature sound and light alarm, and error status alarm |
| Housing | Cold rolling steel with electrostatic spraying exterior; corrosion-resistant chamber platform for long-term laboratory use |
| Shelves | 2 standard shelves; optional shelf quantity up to 7 / 8 / 9 / 13 / 14 depending on model; maximum load per rack 16 kg |
| Permitted load | 45 kg |
| System protection | IP20 according to EN 60529 |
| Voltage | 220–230 V / 50–60 Hz |
| Adjustable exhaust air flap | Inner Ø 28 mm on top for smaller variants; inner Ø 35 mm on top for larger variants |
| Wall clearance | 160 mm back clearance and 100 mm side clearance |
| Test hole | Yes |
| Typical application areas | Microbiology, bacteria cultivation, IVF-related cell culture incubation, pharmacy stability tests, industrial laboratory workflows and medicine research |
| Recovery time after door opening | After 30 s door opening: approximately 8 min at 37°C and 12 min at 50°C |
| Temperature resolution | 0.1°C |
| Temperature uniformity note | Family description states chamber temperature uniformity of 1°C under stated conditions |
| Observation advantage | The inner glass door allows observation of samples without opening the main chamber door and without major temperature loss inside the chamber |
| Accessory / option | Description | Typical note |
|---|---|---|
| Additional shelves | Optional additional shelves according to model capacity | Model-dependent maximum quantity |
| RS485 interface | Communication interface for data integration and monitoring | Optional |
| Printer | Printout option for process data | Optional |
| Program temperature control device | Advanced programmable temperature control option | Optional |
| Remote / WF control | Remote operation and monitoring capability | Optional |
| USB data storage | USB connection and software for acquisition of time and temperature data for operating duration up to 90 days | Optional |
| Touch screen display | Alternative interface upgrade for improved control usability | Optional |
| Temperature certificates | Standard temperature certificate at 40°C in chamber centre issued by COLO technical service; optional one- or multi-point mapping certificate issued by accredited laboratory | Optional accreditation scope depends on request |
| Model | Net / Gross weight | Standard shelves | Optional shelves max |
|---|---|---|---|
| MBIND45 | 27 / 30 kg | 2 | 7 |
| MBIND55 | 30 / 33 kg | 2 | 8 |
| MBIND65 | 32 / 35 kg | 2 | 9 |
| MBIND125 | 45 / 49 kg | 2 | 13 |
| MBIND210 | 58 / 63 kg | 2 | 14 |