COLO.Science · Technical Specification
ALEMADR-DB / DBC · Dry Bath Incubator
| instrument identity | |
|---|---|
| model | ALEMADR-DB (heating) / ALEMADR-DBC (cooling) |
| grade | ≤±0.3°C accuracy @37°C |
| brand / line | COLO.Science · COLO Lab Experts |
| instrument type | Dry bath incubator, semiconductor controlled |
| intended use | Precise heating and cooling for molecular biology, enzymatic reactions, sample incubation, PCR preparation, and cold storage (DBC). |
| temperature control | |
| temp setting range | DB: 0–100°C; DBC: –10–100°C |
| temp control range | DB: RT+5–100°C; DBC: RT–25–100°C |
| time range | 1sec ~ 99min59sec / 1min ~ 99h59min / ∞ |
| accuracy (@37°C) | ≤±0.3°C |
| display accuracy | 0.1°C |
| uniformity (@37°C) | ≤±0.3°C |
| heating / cooling performance | |
| heating time | ≤12 min (25°C → 100°C) |
| cooling time 1 (DBC) | ≤12 min (100°C → 25°C) |
| cooling time 2 (DBC) | ≤25 min (RT → RT–25°C) |
| heating lid | |
| temp setting range | OFF ~ +10°C (relative to block temp) |
| accuracy | ≤±0.5°C |
| heating time | ≤10 min (25°C → 100°C) |
| programming & safety | |
| multi‑point running | Yes (max 5 points) |
| multi‑cycle | Yes (max 99 times) |
| auto running / auto heating | Yes / Yes |
| auto resume after power fail | Yes |
| over‑heating protection | Yes |
| built‑in temperature calibration | Yes |
| storage at 4°C (DBC) | Yes |
| physical & electrical | |
| power | 150 W |
| voltage | AC100 or AC220 V / 50–60 Hz |
| fuse | 250 V, 3 A, Φ5×20 |
| dimensions (mm) | 300 × 200 × 210 |
| weight | DB: 4.0 kg; DBC: 4.7 kg |
| block capacity | see optional blocks |
| LCD screen | Instantly displays operation information. |
| Combined temp control & block | Each block has built‑in sensor and heating/cooling unit for accurate control and faster heating. |
| Interchangeable, auto‑recognized blocks | Enhances flexibility and ease of use. |
| Adjustable heating lid | Allows control of lid temperature (OFF to +10°C). |
| Multi‑point & multi‑cycle modes | Up to 5 points and 99 cycles. |
| Auto restart after power failure | Ensures continuity of experiments. |
| Over‑heating protection | Enhances safety. |
| Built‑in temperature calibration | Maintains long‑term accuracy. |
| Sample incubation in microtubes / microplates |
| Enzymatic reactions and temperature holds |
| PCR preparation steps |
| Heat treatment / conditioning of small volumes |
| Cold storage at 4°C (DBC model) |
| Routine temperature‑controlled protocols in clinical, research, QC labs |
| Code | Capacity | Temp range | Model |
|---|---|---|---|
| PRO4540 | 35×2.0ml centrifuge tube | RT–25…100°C | DBC |
| PRO4541 | 35×1.5ml centrifuge tube | RT–25…100°C | DBC |
| PRO4542 | 54×0.5ml centrifuge tube | RT–25…100°C | DBC |
| PRO4543 | 96×0.2ml centrifuge tube | RT–25…100°C | DBC |
| PRO4544 | 96×0.2ml Elisa plate | RT–25…100°C | DBC |
| PRO4545 | 35×2.0ml centrifuge tube | RT+5…100°C | DB |
| PRO4546 | 35×1.5ml centrifuge tube | RT+5…100°C | DB |
| PRO4547 | 54×0.5ml centrifuge tube | RT+5…100°C | DB |
| PRO4548 | 96×0.2ml centrifuge tube | RT+5…100°C | DB |
| PRO4549 | 96×0.2ml Elisa plate | RT+5…100°C | DB |
*PRO4540–PRO4544 only for cooling model ALEMADR‑DBC.
| topic | recommendation | note |
|---|---|---|
| block selection | Use matching block for your tube/plate type and required temperature range (DB or DBC). | Auto‑recognition simplifies setup. |
| heating lid | Enable lid heating (OFF‑+10°C) to prevent condensation, especially for long runs. | Prevents evaporation and well‑to‑well variation. |
| multi‑step programming | Program up to 5 points with individual hold times; repeat up to 99 cycles. | Ideal for heat‑shock, enzymatic digestions, etc. |
| temperature calibration | Use built‑in calibration to maintain accuracy over time; check with reference thermometer. | Follow user manual procedure. |