COLO.Science | Technical Specification

SIRET-E Series

Analytical Balances | 120-400 g Capacity | 0.1 mg Readability | External Calibration

ANB-SIRET analytical balances are high-precision touch-screen laboratory balances for demanding 0.1 mg measurements. The series combines electromagnetic monoblock force compensation, external calibration with certified E2 control weights, real-time temperature compensation and a three-sided glass windshield for stable analytical weighing.

120-400 g0.1 mg readabilityClass IExternal calibrationDual rangeUSB / RS232Touch screen90 mm pan
Document type: Technical Specification Sheet
Product family: ANB-SIRET Analytical Balances with External Calibration
Models: ANB-SIRET120 / 220 / 320 / 400
Calibration: External calibration with E2 control weight
Series characteristic: 0.1 mg analytical weighing with external calibration and dual-range operation
Format: A4 landscape | COLO.Science TechSpec v2.5
ANB-SIRET analytical balance
ANB-SIRET analytical balanceTouch screen | glass windshield
ANB-SIRET analytical balance | touch screen | external calibration | glass windshield
Capacity range120-400 g
d = 0.1 mg
Address / HQPolje ob Sotli 4
SI-3255, Slovenia
Page 1 / 5
COLO.Science logoSIRET | Analytical Balance Series
Technical data sheet | Informational document
Models: 120 / 220 / 320 / 400 g
SIRET | analytical balance seriesExternal calibration, dual-range weighing and colour touch-screen operation
Max 120-400 gd = 0.1 mgExternal calibrationDual measuring rangeHigh-precision weighing for demanding analytical measurements
Series Overview
Calibration, stability and range control
External calibration: calibration is performed with an external certified calibration/control weight according to the laboratory SOP.
Dual measuring range: automatic dual-range and double-precision function supports efficient handling for small samples.
Temperature compensation: real-time compensation improves analytical stability during routine use.
Anti-vibration filtering: seven-stage adjustable filtering helps adapt the balance to different laboratory environments.
Data handling and usability
PC data workflow: optional communication software can collect data and control the measurement process.
Cursor output: balance measurement data can be written to Excel, Word and other documents following the cursor.
Colour touch screen: direct access to frequent settings with a high-resolution LCD touch display.
Built-in records: date, time and temperature display are adjustable for traceable weighing routines.
Key product features from source PDF
Sensor: electromagnetic monoblock force compensation.
Windshield: glass protective case opens from three sides and reduces air-flow influence.
Interfaces: USB and RS232 communication ports for peripheral equipment.
Units: more than 20 shieldable units including mg, ct, oz and lb.
Operation: automatic mode lock at switch on/off, leveling bubble and portable hand handle.
System: multi-user management, standby mode and external calibration with a certified E2 calibration/control weight.
Core technical summary
ParameterSeries specification
Maximum capacities120 / 220 / 320 / 400 g
Readability0.1 mg; dual-range values listed by model where applicable
Repeatability±0.2 mg
Linearity±0.0003 g
Pan sizeø90 mm stainless steel
DisplayTrue-colour LCD touch screen
InterfacesRS232 / USB
Models4 capacities
Readability0.1 mg
Chamber170 x 160 x 205 mm
Stabilization≤3 s
Page 2 / 5
Models and Technical Specifications
Model selection and source-table values
ModelDual range from sourceMaximum capacityReadabilityCalibrationAccuracy class
ANB-SIRET120120 g / 0.1 mg120 g0.1 mgExternalClass I
ANB-SIRET220220 g / 1 mg and 220 g / 0.1 mg220 g0.1 mgExternalClass I
ANB-SIRET320320 g / 1 mg and 320 g / 0.1 mg320 g0.1 mgExternalClass I
ANB-SIRET400420 g / 1 mg line in source; confirm exact dual-range use400 g0.1 mgExternalClass I
Weighing system and performance
ParameterSpecification
Resolution2,200,000 divisions
USP minimum sample weight400 mg (k = 2, U = 0.1%)
Repeatability±0.2 mg
Reproducibility0.0002 g
Linearity±0.0003 g
Eccentric loading at 1/3 Max0.500 mg
Stabilization time≤3 sec
SensorElectromagnetic monoblock force compensation
CalibrationExternal calibration with certified calibration/control weight
Environment, display and construction
ParameterSpecification
Operating temperature10-30 °C
Relative humidity20-85% RH
Power supply110-230 V, 50-60 Hz by adaptor
DisplayTrue-colour LCD touch screen
InterfaceRS232 / USB
Panø90 mm stainless steel
Windshield175 x 175 x 230 mm (W x D x H)
Measuring chamber170 x 160 x 205 mm (W x D x H)
Housing230 x 310 x 330 mm (W x D x H)
GW / NW7.5 / 5.5 kg
Source note: SIRET-E follows the SIRET external-calibration source specification. Use the source PDF and official offer to confirm any visually compressed dual-range rows. The table above preserves the analytical balance structure and marks the ANB-SIRET400 dual-range row for confirmation before commercial use.
Page 3 / 5
Applications, Delivery, Options and Calibration Weights
SIRET analytical balance
SIRET analytical balanceTouch screen | external calibration
SIRET analytical balance | external calibration | touch screen | 0.1 mg readability
Built-in applications and operating modes
Standard analytical weighing: high-precision weighing for laboratory samples.
Parts counting: counting down to 1 mg in laboratory conditions; 10 mg minimum weight, resolution counting 210,000.
Percent weighing: compare measured mass to a reference.
Density weighing: direct reading function for density workflows with optional hydrostatic set.
Animal weighing: supports unstable sample measurement.
Check and target weighing: upper/lower limit checks, target weighing, summarization, gross/net/tare and peak hold.
Optional accessories
Optional thermal printer
Optional thermal printerMeasurement record output
Optional thermal printer for laboratory balance
OIML E2 calibration weight
OIML E2 calibration weightRecommended analytical balance control weight
Optional OIML E2 calibration / control weight
Ordering note: Source PDF recommends a 200 g E2 class calibration/control weight. Confirm required nominal masses, E2 certificates and calibration/control-weight procedure for the selected SIRET model and laboratory SOP.
Page 4 / 5
COLO.Science | Product Support
Manufacturer and support
COLO Lab Experts
Polje ob Sotli 4, SI-3255, Slovenia
Selection guidance: Send the intended weighing range, sample type, minimum sample mass, dual-range requirement, calibration/control-weight procedure, printer/data-output requirement and documentation expectations.
Quick contact
Web: colo.si
Series: ANB-SIRET120-ANB-SIRET400
COLO.Science | SIRET analytical balance support

Need help selecting a SIRET-E model and E2 calibration-weight configuration?

Send the weighing range, sample type, routine minimum weight, printout requirement, PC data workflow, hydrostatic kit requirement and control-weight procedure. COLO.Science can confirm the suitable SIRET-E model, printer, software and E2 calibration-weight configuration.

Official configuration notice: This document is provided for orientation and product selection. It does not replace the binding specification, delivered configuration, accessory set or acceptance criteria confirmed in the official COLO.Science quotation, proforma invoice, order confirmation or manufacturer-approved technical offer. Confirm model SKU, dual-range configuration where applicable, external calibration procedure, E2 calibration/control masses, certificates, printer, hydrostatic kit and software for the specific order. Specifications may change without notice.