COLO.Science · Product Catalog
VERSAV 50 / 75 · Catalog
| application overview | |
|---|---|
| primary use | Routine laboratory steam sterilization for liquids, instruments, culture media, textiles, rubber goods and waste. |
| sterilization logic | Microprocessor-controlled cycle with preheating, air removal, sterilization holding and drying. |
| air removal | Pulse vacuum exhaust for reliable removal of cold air and improved steam penetration. |
| liquid safety | Selectable exhaust method (quick / low) helps prevent liquid overflow during exhaust. |
| traceability | Built-in printer for recording sterilization data and cycle documentation. |
| program group | examples |
|---|---|
| standard | packed, unpacked, dressings, equipment, rubber |
| liquids | liquid, liquid custom (with liquid probe support) |
| custom | solid custom, culture medium, waste |
| tests | B&D testing, vacuum testing |
| media workflows | sterilization of medium, dissolution of medium |
| safety function | description |
|---|---|
| Over-pressure safety valve | Safety valve opens automatically to release pressure. |
| Over-temperature protection | System powers off automatically if overheating is detected. |
| Door sensor | Door status monitoring; operation stops if door is not secured. |
| Anti-dry-out function | System powers off if water level is too low; error code shown on screen. |
| Thermal safety door cover | Insulated shell reduces burn risk and improves operator safety. |
| filter | 0.22 μm high-efficiency sterilization air filter (filtration efficiency > 99.8%) to keep vacuum-depleted air sterile. |
| parameter | VERSAV50 | VERSAV75 |
|---|---|---|
| total / useful volume | 53 / 50 L | 79 / 75 L |
| chamber size (Ø × H) | Ø400 × 490 mm | Ø400 × 670 mm |
| useful chamber dimension | Ø400 × 420 mm | Ø400 × 600 mm |
| loading height | 690 mm | 870 mm |
| chamber material | Medical stainless steel AISI-316L | |
| design pressure | -0.1 / 0.28 MPa | |
| vacuum system | Eject vacuum system | |
| designed temperature | 150°C | |
| working temperature range | 80 to 144°C | |
| temperature display accuracy | ±0.1°C | |
| sterilization timer range | 0–999 min | |
| delay start | Yes, up to 24 h | |
| circulation patterns | Steam-water internal circulation | |
| display | Touch Screen LCD digital display | |
| water requirement | Distilled / Demineralized | |
| water tank volume | 9 L | |
| dissolution temperature range | 60–100°C (programmable) | |
| temp holding after sterilization | 40–80°C (programmable) | |
| voltage | 220V / 50–60Hz ±10% (16A) | |
| consumption | 4.5 kW | 6 kW |
| external size (W×D×H) | 610 × 700 × 1185 mm | |
| package size | 710 × 800 × 1300 mm | |
| gross weight | 109 kg | 115 kg |
| topic | recommendation | why it matters |
|---|---|---|
| liquid sterilization | Use low exhaust + rapid cooling option for liquid loads; verify with probe when needed. | Helps prevent boil-over / overflow and improves repeatability for validated cycles. |
| load type | Select program based on packaging (packed/unpacked) and material category (rubber, dressings, equipment). | Program parameters align with penetration and drying needs. |
| traceability | Enable printer logging for QC / SOP compliance. | Simplifies audits and batch documentation. |
| utilities | Use distilled/demineralized water and keep filter maintenance in SOP. | Protects steam system and reduces contamination risk. |